Guo Yanyan,Huo Bin,Cheng Junmin,Cao Qiang,Huo Xiaodong,Zhou Xueli,Chai Shude,Wang Haitao,Wang Junjie.The feasibility study of low-dose CT scanning in chest tumor with 125I radioactive seed implantation[J].Chinese Journal of Radiological Medicine and Protection,2017,37(12):950-956
The feasibility study of low-dose CT scanning in chest tumor with 125I radioactive seed implantation
Received:April 06, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.12.014
KeyWords:Low-dose CT  125I radioactive seed implantation  Radiation dose  Image quality
FundProject:国家自然科学基金(81572543);天津市自然科学基金(15JCYBJC28400)
Author NameAffiliationE-mail
Guo Yanyan Department of Radiotherapy, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Huo Bin Department of Oncology, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Cheng Junmin Department of Radiotherapy, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Cao Qiang Department of Oncology, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Huo Xiaodong Department of Oncology, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Zhou Xueli Department of Radiotherapy, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Chai Shude Department of Thoracic Surgery, Second Hospital of Tianjin Medical University, Tianjin 300211, China  
Wang Haitao Department of Oncology, Second Hospital of Tianjin Medical University, Tianjin 300211, China peterrock@126.com 
Wang Junjie Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Hits: 2577
Download times: 1615
Abstract::
      Objective To investigate the feasibility of reducing CT scanning dose in the process of 125I radioactive seed implantation. Methods GEMS phantom and 062 M phantom were scanned using GE Lightspeed RT large hole CT with 120 kV, 100 kV and 80 kV separately, and 150-10 mA (20 mA decreased progressively). The scanning dose, CT value and noise of the region of interest were recorded and the image quality was evaluated. Image signal-to-noise ratio (SNR) and contrast to noise ratio (CNR) values were calculated. Results With the decreasing of tube voltage and current, the SNR values were reduced accorgingly. The values had significant difference with those of standard images except the images acquired with 120 kV, 150-70 mA and 100 kV, 150-90 mA (t=-9.294-3.717, P<0.05). With the decreasing of the tube voltage and current, the CNR decreased significantly. The image quality was too low to evaluate while CNR lower than 2. The high contrast resolution of the CT images were not affected obviously with the tube voltage and current lowering. Conclusions In the process of 125I radioactive seed implantation, it is feasible to choose lower tube voltage and tube current(100 kV, 70 mA) to scan the same area repeatedly except for the first scan with standard chest scanning parameters. The patient radiation dose has fallen dramatically.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9070485  On-line:0

v
Scan QR Code
&et=876B4A42438D29B34F8569CFA947824DEC63262CF53FCA3798488136DDFAE4D5E1B703D27F7E4390479B411F7B36038F27DD00012C92079A6E0B951388B3761C238E145D5A70CF03CF2A1D27EE8185747E8273EA661767B371E886CC8669D56D88977B4073983692834A82C739C49C05&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=58EC03D86D1FF920217446740C798980&vid=&iid=59906B3B2830C2C5&sid=114891522AE71A91&eid=7A7A8B1000B9FF84&fileno=20171214&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="58EC03D86D1FF920217446740C798980";