Wang Zeyang,Niu Shulei,Gao Zhen,et al.The dosimetric effect of different source patterns in case of tumor shrinkage after 125I seed implantation[J].Chinese Journal of Radiological Medicine and Protection,2017,37(9):713-716
The dosimetric effect of different source patterns in case of tumor shrinkage after 125I seed implantation
Received:February 22, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.09.015
KeyWords:125I seed  Treatment planning system  Sources pattern  Dose  Tumor shrinkage
FundProject:
Author NameAffiliationE-mail
Wang Zeyang Section I, Department of Oncology, Hebei Provincial People's Hospital, Shijiazhuang 050000, China  
Niu Shulei Graduate School of Hebei Medical University, Shijiazhuang 050017, China  
Gao Zhen Section I, Department of Oncology, Hebei Provincial People's Hospital, Shijiazhuang 050000, China  
Di Xuemin Graduate School of Hebei Medical University, Shijiazhuang 050017, China  
Du Sui Graduate School of Hebei Medical University, Shijiazhuang 050017, China  
Zhang Hongtao Section I, Department of Oncology, Hebei Provincial People's Hospital, Shijiazhuang 050000, China  
Wang Juan Section I, Department of Oncology, Hebei Provincial People's Hospital, Shijiazhuang 050000, China niushulei167x@163.com 
Hits: 3467
Download times: 1992
Abstract::
      Objective To study the dosimetric effect of two source patterns, including equal spacing and peripheral dense intermediate sparse by assuming a tumor shrinking speed of 20% per month after 125I seed implantation.Methods A virtual cylindrical tumor with 4 cm in height and 5 cm in diameter was contoured on a three-dimensional treatment planning system (TPS). Two groups of preoperative plans were made with 1.85×107 Bq 125I seeds using two source patterns respectively. It was assumed that the tumor height was unchanged, while the diameter of tumor would decrease at a speed of 20% per month, and the locations of seeds would concentrate towards the tumor core. The 90% target volume dose (D90), the ratio of 90% isodose volume over the target volume (V90), and the ratio of 150% isodose volume over the target volume (V150) were calculated at 0, 1, 2, 3 months after 125I implantation respectively.Results In equal spacing group, 85 seeds were implanted. The values of D90 were 126.20, 130.41, 133.82 and 139.48 Gy after 0, 1, 2 and 3 months respectively. The values of V90 were 97.0%, 98.1%, 99.3% and 100%, while those of V150 were 70.2%, 69.9%, 71.1% and 71.5%. The dense in-periphery and sparse-in the middle group was loaded with 75 seeds. The D90 values were 126.46, 125.41, 123.50 and 128.83 Gy, the V90 95.2%, 95.7%, 94.9% and 97.6%, and the V150 52.8%, 60.4%, 62.7% and 59.3% after 0, 1, 2 and 3 months, respectiviely.Conclusions When the tumor diameter reduces at a rate of 20% per month after 125I seed implantation, the expected tumor dose absorption will gradually increase using the equal spacing sources pattern. However, the expected dose does not vary with source distribution of dense-in the-surrounding and sparse-in-middle, which also reduces high dose volume more than the equal spacing pattern.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029370  On-line:0

v
Scan QR Code
&et=E29B04634B0E836E41D9D7BC52B55AA265BF70146B8A07F10FF87567DC669EB96A40BBDB51216977F23EA857012ABFD8E0E53C6D954EBDD6D7B264172A0CBBCD9CDFE1C0A27E395DB73F2DD4F81FE6C42B9317E903AC62BF17085150FF9493BDC499F1F33A969095D8E995BF292DF997D4971C87FF1E2B62&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=CF1332E5A7C16DE7B4C2F35708A86C64&vid=&iid=9CF7A0430CBB2DFD&sid=A586B761C9AA2FAA&eid=27350781F1397F32&fileno=20170916&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="CF1332E5A7C16DE7B4C2F35708A86C64";