Wang Qiang,Xiao Jianghong,Fu Yuchuan,Li Tao,Li Xia,Song Ying,Bai Sen.Script-based automated placement of gantry angles for tangential beams of breast cancer[J].Chinese Journal of Radiological Medicine and Protection,2017,37(9):677-680
Script-based automated placement of gantry angles for tangential beams of breast cancer
Received:June 13, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.09.007
KeyWords:Breast cancer  Radiotherapy  Tangential beams  Script  Penalty function
FundProject:科技部基础性工作专项资金(2014FY211000)
Author NameAffiliationE-mail
Wang Qiang Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China  
Xiao Jianghong Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China  
Fu Yuchuan Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China  
Li Tao Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China  
Li Xia Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China  
Song Ying Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China  
Bai Sen Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China baisen@scu.edu.cn 
Hits: 2935
Download times: 2091
Abstract::
      Objective To reduce the dose to organ at risk (OAR) by script-based optimizing the gantry angles of tangential beams for forward-planned intensity-modulated breast radiotherapy (fIMRT) plans.Methods A total of 46 patients with breast cancer (26 in left and 20 in right) after conservative surgery who were treated with manually planned fIMRT were retrospectively re-planned using a script to place the tangential beams automatically. The script program was developed using Pinnacle3 scripting language as imbedded in the Pinnacle3 treatment planning system, and the Python scripting language. Dosimetric indices, including V20 of the ipsilateral lung, D1 of the contralateral breast and V10 of the heart were extracted by the script for plans of different gantry angles. Then the optimal tangential angles were selected by a penalty function.Results For the 26 left breast cancer cases, V20 of the ipsilateral lung for manual and script were (18.2±3.4)% and (18.1±3.4)%, respectively(P>0.05), while D1 of the contra-lateral breast(249.2±88.4) cGy and (230.4±80.2) cGy, respectively( t=2.27, P<0.05), V10 of the heart (13.7±5.7)% and (13.1±5.7)%(t=2.82, P<0.05). The V20 of the ipsilateral lung of the 20 right breast cancer cases were (19.8±3.8)% and (19.8±3.9)%, respectively(P>0.05). The script plans achieved better V10 of the heart than the manual plans (t=2.53, P<0.05).Conclusions Compared with the manualmethod, the script can assess of the impact of tangential beam angles on the OAR dose, patient-specifically and quantitatively, hence make better trade-offs and improve the plan quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9079267  On-line:0

v
Scan QR Code
&et=6B1A703BF7FDE086CEAFB272B97145BDAC34E546A594FD538F1293F94FD0DCDEC93A69BD64DC205E69CA801531DCFD088D1555C6B130BE6C19E763C4627D583BBD6FCCECFD38941A2E7B3001AEAFAC74779EB2DFBC804B41&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=CF1332E5A7C16DE799F6A6383A49E5FC&vid=&iid=9CF7A0430CBB2DFD&sid=2E4E3741E8FB64E9&eid=34D13857B558254E&fileno=20170908&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="CF1332E5A7C16DE799F6A6383A49E5FC";