Li Mingsheng,Jiang Qinghuan,Liu Ya,Cheng Jinsheng.Study on step accuracy measurement and dose deviation of 252Cf neutron afterloading radiotherapy machine[J].Chinese Journal of Radiological Medicine and Protection,2017,37(8):626-629
Study on step accuracy measurement and dose deviation of 252Cf neutron afterloading radiotherapy machine
Received:January 18, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.08.014
KeyWords:EBT3 film  Double ion chamber method  252Cf neutron source
FundProject:
Author NameAffiliationE-mail
Li Mingsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevetion, Bejing 100088, China  
Jiang Qinghuan Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevetion, Bejing 100088, China  
Liu Ya Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevetion, Bejing 100088, China  
Cheng Jinsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevetion, Bejing 100088, China addila@qq.com 
Hits: 2522
Download times: 1555
Abstract::
      Objective To study the step accuracy of 252Cf neutron afterloading radiotherapy machine and the dose deviation caused by step deviation. Methods EBT3 film was used to measure the steps of the 252Cf neutron source, and then the center of each 252Cf neutron source was identified by measuring the optical density value by using the ImageJ software. Double ion chambersmethod was used to measure the dose deviation due to the 252Cf neutron source position shift. Results 252Cf neutron source step accuracy may amount to 0.01 mm using EBT3 film measurement, when 252Cf neutron source position deviation is less than 3 mm, the dose deviation is less than 2.5%. Conclusions The study on the step accuracy and position deviation of the 252Cf neutron source can provide a reference for the quality control standard of the 252Cf neutron afterloading radiotherapy machine.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9057663  On-line:0

v
Scan QR Code
&et=08EE5A5866ACE44078B8FD1BCD9A87FEA99EFCE42E9E8CC6F67F27ED2A543FE797BA3C14BAAF3740466BD95FA5F4274249BE09536489778701FE77BACA122EC8DDC4A79E8E8434ED5395CAA3CCFB2E4F063482BC5FF20D7012372A94FF17BAEE08E3A1C0EFF7E51FA070884DCE86871FFBF1EEFA7A92D86A&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=3C3EF3B75913B4FF62C9DD616716C331&vid=&iid=5D311CA918CA9A03&sid=C824C8F9F54AE9B9&eid=E4EC39E73004B593&fileno=20170814&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="3C3EF3B75913B4FF62C9DD616716C331";