Luo Wei,Zhao Yue,Wang Jianchao.Eatimation of effective doses from a dental cone beam computed tomographic scanner[J].Chinese Journal of Radiological Medicine and Protection,2017,37(3):222-225
Eatimation of effective doses from a dental cone beam computed tomographic scanner
Received:June 22, 2016  
DOI:10.3760/cma.j.issn.0254-5098.2017.03.012
KeyWords:Cone beam computed tomography(CBCT)  Dose-area product(DAP)  Effective dose  Conversion coefficient
FundProject:
Author NameAffiliationE-mail
Luo Wei Department of Dental, People's Hospital of Peking University, Beijing 100044, China  
Zhao Yue Department of Dental, People's Hospital of Peking University, Beijing 100044, China  
Wang Jianchao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radialogical Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China wjc5188@163.com 
Hits: 3202
Download times: 1900
Abstract::
      Objective To calculate effective dose based on the dose-area Product (DAP) measured in dental cone beam computed tomograhic (CBCT) scanning and using a conversion coefficient. Methods The method of the DAP was used to calculate the effective dose from KODAK 9500 dental CBCT scanner with fields of view (FOV) of 16 cm×15 cm and 12 cm×8 cm. The DAP-values (mGy·cm2) were measured using a transmission ionization chamber connected to an electrometer (Doseguard 100, RTI Electronics AB, Sweden) and the chamber was placed directly on the exit of the X-ray tube. All measurements were repeated three times and averaged on each FOV data. The conversion coefficient of 0.080 μSv/mGy·cm2 was used to estimate the effective dose on the basis of DAP values, which was quoted from a literature. Results DAP values were found between the two FAOs to vary from 512 to 1 083 mGy·cm2, and the effective doses calculated with 0.080 μSv/mGy·cm2 were 41 to 87 μSv. The effective doses given in this paper was lower than those reported in other literatures. Conclusions DAP measurement was found to be well defined and easy-to-use method of determining effective dose for some CBCT Units. However, determination of specific conversion coefficient in the CBCT must be further developed.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9062549  On-line:0

v
Scan QR Code
&et=A854F17E28E369E296F930567B8C4A2B2A7BB54B5DCB8D143A0E835077F904320B7C2191DCCF77AEE9FF86B296ACEDC9CF6EF88A28CF8B6F4B1E2371B5839D29E63CC6D827319AC45F55BDD964A37C611FA05E431C5B2B64&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=85B39D5F58480BE3CA507F90D3C66B14&vid=&iid=38B194292C032A66&sid=8B59EA573021D671&eid=4966445AEEBA9556&fileno=20170312&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="85B39D5F58480BE3CA507F90D3C66B14";