Li Chengqiang,Tao Cheng,Zhu Jian,Lu Jie,Liu Tonghai,Yin Yong.Analysis of multi-leaf collimator leaf position deviation in volumetric modulated arc therapy[J].Chinese Journal of Radiological Medicine and Protection,2016,36(12):904-908
Analysis of multi-leaf collimator leaf position deviation in volumetric modulated arc therapy
Received:July 04, 2016  
DOI:10.3760/cma.j.issn.0254-5098.2016.12.005
KeyWords:Volumetric modulated arc therapy  Dynalog file  Multi-leaf collimator  Position error
FundProject:国家自然科学基金(81301298,81472811);山东省自然科学基金(ZR2011HL060)
Author NameAffiliationE-mail
Li Chengqiang Shandong Cancer Hospital, Jinan 250117, China  
Tao Cheng Shandong Cancer Hospital, Jinan 250117, China  
Zhu Jian Shandong Cancer Hospital, Jinan 250117, China  
Lu Jie Shandong Cancer Hospital, Jinan 250117, China  
Liu Tonghai Shandong Cancer Hospital, Jinan 250117, China  
Yin Yong Shandong Cancer Hospital, Jinan 250117, China yongyinsd@163.com 
Hits: 2844
Download times: 1783
Abstract::
      Objective To evaluate the position deviation of multi-leaf collimator (MLC) in volumetric modulated arc therapy for nasopharyngeal carcinoma and to verify a method of detecting the leaf position accuracy. Methods Thirty-two volumetric modulated arc plans of nasopharyngeal carcinoma were randomly selected. A clockwise arc with a gantry range of 181° to 179° and a counterclockwise arc with a gantry range of 179° to 181° were used for each plan. Each arc contained two leaf groups, groups A and B. By using the Argus software, the actual MLC leaf positions were extracted from the Dynalog files in the Varian Trilogy accelerator. The planed MLC leaf positions were compared with the actual positions, and the differences were calculated between them. Results The proportions of the leaf position errors within ±2, ±1.5, ±1 and ±0.5 mm were 99.99%, 99.90%, 99.07% and 93.98%, respectively. The proportions of the leaf position errors within ±1 mm were 98.08%, 100%, 98.97% and 99.01% for the first therapeutic arc (ARC1), second therapeutic arc (ARC2), A and B groups, respectively. The leaf position errors range of ARC1 and ARC2 were -2.95 to 2.99 and -0.22 to 0.23 mm (t=2.35, P<0.01). The leaf position errors range of leaf bank A and leaf bank B were -2.95 to 2.68 and -2.92 to 2.99 mm (P>0.05). Conclusions During the volumetric modulated arc radiotherapy, the leaf positions deviation are in control. The Dynalog files can be used as an effective way to analysis the leaf position errors.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9058888  On-line:0

v
Scan QR Code
&et=6FD48CA0A0DD47594153DA0B8F60CB050E5574BE42F0F6761A27B8A7D73FDCB666B337C4A293FCC00D0A64C4F2759D4469AF43151196C2FA6B73A487E17E2E6658D6BBC7EDB4E0A899F277A56CF5772432AE223543EB6B8CAD8202A71CBF9855&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=1B4F2A2B6C1A4B2AE2E76553D57E3F7A&vid=&iid=59906B3B2830C2C5&sid=46CB56AABC2765FF&eid=51FE5F6F4794AD16&fileno=20161205&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="1B4F2A2B6C1A4B2AE2E76553D57E3F7A";