Zhu Chaohua,Wei Chunjiao,Xia Bangchuan,Yu Zhiyong,Xu Zihai,Chen Chaomin.Development of nasopharyngeal carcinoma radiotherapy positioning of CT/MRI image registration auxiliary device[J].Chinese Journal of Radiological Medicine and Protection,2016,36(11):832-836
Development of nasopharyngeal carcinoma radiotherapy positioning of CT/MRI image registration auxiliary device
Received:September 18, 2016  
DOI:10.3760/cma.j.issn.0254-5098.2016.11.007
KeyWords:Radiation therapy  Nasopharyngeal carcinoma positioning  CT/MRI  Image registration  Image fusion
FundProject:广东省科技计划项目(2015A020214013)
Author NameAffiliationE-mail
Zhu Chaohua Radiotherapy Department, No. 303 Hospital of PLA of Nanning 530021, China  
Wei Chunjiao Radiotherapy Department, No. 303 Hospital of PLA of Nanning 530021, China  
Xia Bangchuan Radiotherapy Department, No. 303 Hospital of PLA of Nanning 530021, China  
Yu Zhiyong Radiotherapy Department, No. 303 Hospital of PLA of Nanning 530021, China  
Xu Zihai Radiotherapy Department, No. 303 Hospital of PLA of Nanning 530021, China  
Chen Chaomin Medical Instrument Research Institute, School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China gzccm@fimmu.com 
Hits: 2658
Download times: 2339
Abstract::
      Objective Developing a CT/MRI image registration assistant device for radiotherapy localization of nasopharyngeal carcinoma, to achieve precise fusion, and by analyzing the differences of the outline to test the application of the registration and fusion image. Methods Design a specific shape pillow for the head-neck coil of MRI, which could keep the patient have the same position form for CT scan. Using paired t-test to analyze the difference of the parotid gland and the mandible. Results After adopting the new device, the registration accuracy has a significant improvement for the head and neck CT and MRI image registration. Comparing two sets of image outline volume, the outlined volume of parotid gland in CT image is smaller than that of CT/MRI fusion image Left parotid gland volume comparison:fusion (17.78±6.89) cm3 and CT(17.17±7.02) cm3(t=-2.715, P<0.05). Right parotid gland volume was (19.23±8.91) cm3 for fusion, and (17.47±7.42) cm3 CT for(t=-2.552, P<0.05), and the size of the mandible was larger than that of CT/MRI fusion image.Left mandible volume was (33.7±5.59) cm3 for fusion, and (34.8±6.27) cm3 for CT (t=3.548, P<0.05). Right mandible volume was (34.46±6.08) cm3 for fusion, and (35.38±6.72) cm3 for CT(t=3.14, P<0.05). Conclusions The accurate image fusion of head and neck CT and, MRI could be achieved with this device, which might be helpful for the delineation of clinical target volume.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9068154  On-line:0

v
Scan QR Code
&et=A90EAC7419FB9B220ECBF6D77DD06AD225973C47EF5480DE8A9083C8C5C7A7CEBB74BA579A98D7349E0CA0930DA707A0804F9DF7D2645653AEA72F4F1BC114A7BC2FDCC9F720733D46A091F6814D21D0CC2A2E67CD9E6DD57F22BF71381DB3EA6BE9E0BABB31F17FC6D6053D19C7B618&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=2CB0EE8C6BD1AB59EAB73A238DAD0647&vid=&iid=708DD6B15D2464E8&sid=AB1DE136C335A86C&eid=BC60A9A1D91963F5&fileno=20161107&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="2CB0EE8C6BD1AB59EAB73A238DAD0647";