Yuan Jilong,Li Mingsheng,Cheng Jinsheng.Quality control research of active spot scanning proton and heavy ion accelerator's radiation field uniformity[J].Chinese Journal of Radiological Medicine and Protection,2016,36(8):626-629
Quality control research of active spot scanning proton and heavy ion accelerator's radiation field uniformity
Received:September 29, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2016.08.014
KeyWords:Proton and heavy ion accelerator  Active spot scanning  Radiation field uniformity
FundProject:国际原子能机构国际合作项目(CPR6006)
Author NameAffiliationE-mail
Yuan Jilong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Mingsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Cheng Jinsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China chengjs3393@163.com 
Hits: 2379
Download times: 2267
Abstract::
      Objective To develop the method for testing the consistency of irradiation field produced by the active spot scanning proton and heavy ion accelerator. Methods Calibration of the EBT3 films were carried out with the calibrated ion beam to establish the dose calibration curve. According to the different proton and carbon ion energies (proton:94.29, 150.68, 212.62 MeV; carbon ion:175.99, 283.43, 412.54 MeV/u), EBT3 films were located in the solid water phantoms in each therapy room, respectively. Finally, the irradiated EBT3 films were scanned and the radiation field size's deviation and flatness were analyzed. Results In different conditions, radiation field size's deviations were all less than 2 mm and the flatness parameters were all controlled below the 5%. Conclusions EBT3 films can be used to test the active spot scanning proton and heavy ion accelerator's radiation field uniformity.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9071888  On-line:0

v
Scan QR Code
&et=F27DD3FEE5F2566CAFEA97DA18ED9D19797C014A6FAB3FBE097E6685E4FA9546A9A01DFDE4C370489B369FDDA9906E05CF89AF7BA562F291154B8AB5AA2DF29515B0784FF25EB705D406616823F40E4DA89C93CF4E657A56FB11B2C7F9E63F7AB0BDF5984E16CDBC6704AA5D3EAB0FF1&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=B25A86971E857B044D8A25F1B7045C46&vid=&iid=5D311CA918CA9A03&sid=C824C8F9F54AE9B9&eid=E4EC39E73004B593&fileno=20160814&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="B25A86971E857B044D8A25F1B7045C46";