Cheng Jinsheng,Yuan Jilong,Li Mingsheng.Research on quality testing for active spot scanning proton and heavy ion accelerator[J].Chinese Journal of Radiological Medicine and Protection,2016,36(8):621-625
Research on quality testing for active spot scanning proton and heavy ion accelerator
Received:September 29, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2016.08.014
KeyWords:Proton and heavy ion accelerator  Active spot scanning  Quality test
FundProject:国际原子能机构国际合作项目(CPR6006)
Author NameAffiliation
Cheng Jinsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Yuan Jilong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Li Mingsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 2377
Download times: 3198
Abstract::
      Objective To study the quality testing of dose delivery system of the active spot scanning proton and heavy ion accelerator, in order to provide the reference for the quality control of related equipment.Methods In the four therapy rooms, both 0.6 cc chambers and Gafchromic EBT3 films were used, respectively, to test the accelerator for dose reproducibility, dose linearity, dose stability, depth dose distribution, beam scanning position deviation and radiation field uniformity in each therapy room. Results Dose reproducibility variation coefficients are all less than 1.5%, dose linearity's maximum deviations less than 2%, dose stability's deviations less than 2%, depth dose distribution stability within 2%, beam scanning position deviation less than 1 mm, consistency of irradiation field's deviation less than 2 mm, and flatness within ±5%. Conclusions The indicators about quality testing for the active spot scanning proton and heavy ion accelerator are all in line with the requirements of IEC standards draft.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9078728  On-line:0

v
Scan QR Code
&et=FC9DA7BF2F842A3B1EEE1BDFAD9A1DC946EDA0DF0B372219FCDEEC7DEF69065D54313B5A6C01E7552312AF99D3536A1322C8FEB4C746BE4690BAA5F51B008A73AC4BBBDB5420DAA2E2112DFE5C2DC3DE5B6ADBC2C72D41CE&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=B25A86971E857B043BBBE73C299B64F0&vid=&iid=5D311CA918CA9A03&sid=BA3451F2C9E4FB70&eid=42FF82ADD37D41AA&fileno=20160813&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="B25A86971E857B043BBBE73C299B64F0";