Ge Ning,Han Dongliang,Yi Fengtao,Gu Shiyong,Chen Fuci,Wang Zhibin.Significance of radiological protection of sensitive organs in non-treated region in body γ-knife treatment[J].Chinese Journal of Radiological Medicine and Protection,2016,36(6):456-460
Significance of radiological protection of sensitive organs in non-treated region in body γ-knife treatment
Received:January 21, 2016  
DOI:10.3760/cma.j.issn.0254-5098.2016.06.012
KeyWords:Body γ-knife  Lead shielding  Sensitive organs  Radiation protection
FundProject:
Author NameAffiliation
Ge Ning Department of Radiotherapy, Wuhan General Hospital of Guangzhou Command, Wuhan 430070, China 
Han Dongliang Department of Radiotherapy, Wuhan General Hospital of Guangzhou Command, Wuhan 430070, China 
Yi Fengtao Department of Radiotherapy, Wuhan General Hospital of Guangzhou Command, Wuhan 430070, China 
Gu Shiyong Department of Radiotherapy, Wuhan General Hospital of Guangzhou Command, Wuhan 430070, China 
Chen Fuci Department of Radiotherapy, Wuhan General Hospital of Guangzhou Command, Wuhan 430070, China 
Wang Zhibin Center for Disease Control & Prevention of Guangzhou Command, Guangzhou 510000, China 
Hits: 2309
Download times: 1991
Abstract::
      Objective To explore the radiation dose from body γ-knife treatment to the non-targeted region's sensitive organs before and after shielding. Methods 20 patients suffering tumors less than 5 cm were selected. Calibrated thermoluminescence dosimeters (TLD) were placed above the sensitive organs, such as lens, thyroid gland and sexual gland, to measure the radiation dose received before shielding. Different plans were prepared for the patients with lung and renicapsule tumors using calibrators of different size. Radiation dose was measured by the aid of water phantom. For selected lung-tumor treatment plan, the radiation doses were measured at the same location on the water phantom shielded with and without 1, 2 and 4 cm lead, respectively. Results The maximum doses were 1 023.3 mGy for lens, 1 235.7 mGy for thyroid gland and 1 176.8 mGy for sexual glands after treatment, respectively, being higher on the left site than the right. The radiation doses to the sensitive organs were higher for the water phatom with more tartgted points, decreasing by 55%-91% after being covered with 1, 2 and 4 cm lead shieldings. There were significant differences in doses received before and after lead shielding (t=14.4, 12.9, 13.3, P<0.05). Conclusions In the course of body γ-knife treatment, the additional factors would increase the dose to the sensitive organs. Therefore, it is necessary to provide lead shielding protection to the teenagers and adults with fertility when they undergo body γ-knife treatment. Trial registration Chinese Clinical Trial Registry, ChiCTR-OOC-16008259.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9060542  On-line:0

v
Scan QR Code
&et=331551E9D38B96EB3A04A27BE4A2F70A12C0888DF8DED47040B5F274D80EEE301858230644FEB03B86DE8337F832823E819B374AA27B41C51F2D4F66B593620BF876614EA639D393E8F595A8A3207E764257A0907401554C89D5E2FDCBC955F45B290B9FF71A2E9AB458AC7B46B0A4A6&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=D3E0C87FB6A41B569FD8A9FE304D1574&vid=&iid=B31275AF3241DB2D&sid=522844664D9E629A&eid=E0172F1A638CE984&fileno=20160612&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="D3E0C87FB6A41B569FD8A9FE304D1574";