Yang Jing,Chen Mi,Liang Zhiwen,et al.Evaluation on repeatability of the aperture sizes of CyberKnife VSITM Iris collimator[J].Chinese Journal of Radiological Medicine and Protection,2016,36(4):291-293
Evaluation on repeatability of the aperture sizes of CyberKnife VSITM Iris collimator
Received:December 09, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2016.04.012
KeyWords:CyberKnife  Iris variable aperture collimator  Film
FundProject:
Author NameAffiliationE-mail
Yang Jing Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China  
Chen Mi Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China  
Liang Zhiwen Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China  
Hu Bin Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China  
Cheng Junping Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China  
Li Qin Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China 675181445@qq.com 
Hits: 3701
Download times: 2580
Abstract::
      Objective To evaluate the repeatability of the aperture sizes based on the Iris variable aperture collimator (Iris collimator) of CyberKnife VSITM. Methods Radiochromic films were placed at 800 mm of source-to-axis distance (SAD), and Iris quality assurance (QA) software was used to extract the information from the digital images, which were exposured by X-ray beams. Results The average aperture accuracy of Iris collimator was better than 0.11 mm over the beam range 5-40 mm, and the accuracy for 50 mm and 60 mm beams was within 0.19 mm. Conclusions The results suggest a good repeatability of aperture sizes for the CyberKnife VSITM Iris collimator.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11565270  On-line:0

v
Scan QR Code
&et=D700F206953109B386D35682B4C5EA41067DCF75502C8D139800F4ECAD1F27DD95D2370B445E41575424F3864EC7EDBD5B133A5ECDBF3A02AC414CDD0A5CC87BCDA884EA7AE8F26E32A5EF5A6F0E8486E8FA922AF5EF05185C14C9148732D427BA003BB8A09B5096&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=28B2676AB082C1DC4B94E0EA60C7E1F3&vid=&iid=E158A972A605785F&sid=6490F0E20C4B41AD&eid=2AC7DCCBBC26ECF8&fileno=20160412&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="28B2676AB082C1DC4B94E0EA60C7E1F3";