Liu Hui,An Jinggang,Song Ying.SPECT system performance test and analysis of results[J].Chinese Journal of Radiological Medicine and Protection,2015,35(10):767-770
SPECT system performance test and analysis of results
Received:September 19, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.10.011
KeyWords:SPECT  NEMA standard  Resolution  System planar sensitivity
FundProject:公益性行业科研专项(201002009)
Author NameAffiliation
Liu Hui Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
An Jinggang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Song Ying Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 2546
Download times: 2982
Abstract::
      Objective To analyze the influence factors on, and the relationship between, the system performance parameters by testing SPECT equipment. Methods By reference to National Electrical Manufactures Association standards and manufacturer's specifications, the performances of a total of 31 SPECTs in 12 provinces were measured, for the first time, with regard to their system spatial resolution (SSR), system planar sensitivity (SPS) and tomographic spatial resolution (TPR). Results The results were as follows: (7.90±0.62) mm for SSR, with the highest 9.46 mm and the lowest 7.04 mm; (78.54±13.17)s-1·MBq-1 for SPS, with the highest 123.80 s-1·MBq-1 and the lowest 56.70 s-1·MBq-1; and (13.12±2.59) mm for TSR, with the highest 18.13 mm and the lowest 8.45 mm. These values indicated a nearly consistent upward and downward trend and could meet the clinical requirements by comparison with the manufactuer's specifications. Conclusions Mutual restrictions have been shown between SSR and SPS. Increased thickness of the system crystal plane can improve the SPS, but also has a negative impact on the SSR. For all the above reasons, it is the optimum solution to choose the right type of collimator and crystal thickness for different clinical applications.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9077013  On-line:0

v
Scan QR Code
&et=B8DAAE961D60951965F3337DD9D95878E66A88E22BC5A2F3BB484E3AF31D1749C56B4851CACD8EF591C87AE1FE435B35D6DDD81B072C22AB&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=6F3C2608457E70065A576D55BB9E2CF1&vid=&iid=F3090AE9B60B7ED1&sid=70081086C74F4110&eid=BDEE8BA20F4733DB&fileno=20151011&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="6F3C2608457E70065A576D55BB9E2CF1";