Liu Yunfu,Zhang Zongrui,Kang Tianliang,Ma Wentao,Niu Yantao.A phantom study of the influence of tube voltage on image quality and radiation dose in DR chest radiography[J].Chinese Journal of Radiological Medicine and Protection,2015,35(9):713-716
A phantom study of the influence of tube voltage on image quality and radiation dose in DR chest radiography
Received:April 08, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2015.09.019
KeyWords:kV value  Effective dose  Relative noise  Contrast-to-noise
FundProject:
Author NameAffiliationE-mail
Liu Yunfu Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Zhang Zongrui Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Kang Tianliang Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Ma Wentao Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Niu Yantao Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China ytniu163@163.com 
Hits: 3745
Download times: 2838
Abstract::
      Objective To investigate the influence of tube voltage on image quality and radiation dosage to subjects in chest digital radiography (DR). Methods An adult chest phantom was exposed in the range of 80-130 kV with the interval of 10 kV and the automatic exposure control (AEC) setting was selected from -4 to 4. The entrance skin dose (ESD) of the phantom was measured, and the relative noise, contrast-to-noise ratio (CNR) and the effective dose of each exposure was calculated. Results The ESD was (0.062 9±0.027 4) mGy, and the effective dose was (0.012 7±0.004 5) mSv. The effective dose increased linearly with the ESD, and there was a positive correlation (r=0.912, P<0.01). When the tube voltage was held constant, the relative noise decreased as the effective dose increased (r=-0.967, -0.969, -0.968, -0.969, -0.968, -0.970, P<0.01). The CNR increased with the effective dose increasing under the same tube voltage (r=0.987, 0.987, 0.986, 0.987, 0.988, 0.989, P<0.01). The ESD and effective dose decreased as the kV values increased under the same AEC, and the degree of decrease could be up to 50% and 20%. The relative noise decreased as the kV values increased under the same AEC, and the degree of decrease could be up to 23%. The CNR values increased up to 8%. Conclusions The use of higher kVp in chest digital radiography can reduce the radiation dose to some extent.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9067927  On-line:0

v
Scan QR Code
&et=5128A264A75270F483CC666099D83C61417D0F9C1E5C020B8E91608E4BD0451BBCCC84B5AC460175D1529A7B34FCB771DBDC557A4215F02D370BA00BC16A093420304676F59A26AE21E8B375EBAC2061AA56C03FA9CC48D54C18619281501E5F6337CEB575EFD6C48E28BA00BF78AC5C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=B388734F3248585CE7F3A2072E48204B&vid=&iid=9CF7A0430CBB2DFD&sid=A586B761C9AA2FAA&eid=27350781F1397F32&fileno=20150919&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="B388734F3248585CE7F3A2072E48204B";