Zhao Nan,Yang Ruijie,Wang Junjie.The dosimetric characteristics of thermoluminescent dosimeter[J].Chinese Journal of Radiological Medicine and Protection,2015,35(9):696-699
The dosimetric characteristics of thermoluminescent dosimeter
Received:November 17, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.09.014
KeyWords:Thermoluminescent dosimeter  Repeatability  Dose response  Dose rate effect  Energy response
FundProject:国家自然科学基金(81372420)
Author NameAffiliationE-mail
Zhao Nan Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Yang Ruijie Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China ruijyang@yahoo.com 
Wang Junjie Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Hits: 3139
Download times: 2673
Abstract::
      Objective To study the dosimetric characteristics of thermoluminescent dosimeter (TLD). Methods A total of 350 pieces of TLD 2000 type TLDs having the same sensitivity within±3.0% were selected. The TLDs were irradiated to study the repeatability and dosage effect of TLDs by using 6 MV X-rays and 125I seeds. In order to study the dose response, the TLDs were irradiated at different dosages by using 137Cs (662 keV γ-rays), 125I seeds and 6 MV X-rays. In order to study the energy response, the TLDs were exposed to the same dose, from 125I seeds, 137Cs and X-rays (48, 65, 83, 118, 250 keV and 6 MV). Results The maximum deviations of the repeatability were 2.7% and 4.0% for 6 MV X-rays and 125I seeds, respectively, and there was no effect of dose rate observed. The dose response of TLDs to 137Cs and 125I seeds were linear. For 6 MV X-rays, the linear response was within the range of 0.74-10 Gy and non-linear range was beyond 10 Gy. The energy response to 125I seeds, 48, 65, 83, 118, 250 keV and 6 MV X-rays, relative to the energy response of 137Cs, were 1.70, 1.25, 1.08, 0.99, 0.91, 0.96 and 1.22, respectively. Conclusions TLD 2000 has a good repeatability and linear dose response for 137Cs, 125I seeds and 6 MV X-rays with no dose rate effects, but the dose response is energy dependent.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061993  On-line:0

v
Scan QR Code
&et=E29B04634B0E836E386CF964BF8E02ABD7AB803BFA720646C2E059B9F2470DC8B1CC2EC6D8FA2E96BD852452AA2FF25136E28AD77EEBEBC5BEAD0261BB2FDA0F&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=B388734F3248585CA09EBC324119272C&vid=&iid=9CF7A0430CBB2DFD&sid=A7379F6713A46835&eid=C230D48775108CE3&fileno=20150914&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="B388734F3248585CA09EBC324119272C";