Cao Guoquan,Cai Zhimin,Chen Yongchun,et al.Influence of low tube voltage in combination with simultaneous iterative reconstruction algorithm on image quality and radiation dose in thoracic CT angiography:a chest phantom study[J].Chinese Journal of Radiological Medicine and Protection,2015,35(5):393-396
Influence of low tube voltage in combination with simultaneous iterative reconstruction algorithm on image quality and radiation dose in thoracic CT angiography:a chest phantom study
Received:October 09, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.05.020
KeyWords:Tomography, X-ray computed  Tube voltage  Iterative reconstruction  Radiation dose  Thoracic CT angiography
FundProject:浙江省公益性技术应用研究计划(2011C23021);温州市科技局科技计划项目(Y20130146)
Author NameAffiliationE-mail
Cao Guoquan 325000 温州医科大学附属第一医院放射科  
Cai Zhimin 325000 温州医科大学附属第一医院医学工程处  
Chen Yongchun 325000 温州医科大学附属第一医院放射科  
Chen Xiao 325000 温州医科大学附属第一医院放射科  
Tai Yunpeng 325000 温州医科大学附属第一医院放射科  
Pan Kehua 325000 温州医科大学附属第一医院放射科  
Chen Weijian 325000 温州医科大学附属第一医院放射科 wyyycwj@163.com 
Wang Zhenzhang 325000 温州医科大学附属第一医院放射科  
Hits: 4309
Download times: 3252
Abstract::
      Objective To study the effect of low-tube voltage combined with the iterative algorithm(IR) on image quality and radiation dose in thoracic CT angiography. Methods A simulated blood vessel of chest was constructed using an ATOM 701-D adult male chest phantom and iodine-water mixture contrast medium. A 320-row volume CT was used to scan the chest at tube voltages of 80, 100 and 120 kV in combination with off, mild, standard and strong IR levels (12 protocols in total), respectively. Tube currents and other parameters were not changed during scanning. The radiation doses were recorded. CT value of the transverse infusion tube and its corresponding noise, CT value of mediastinal soft tissues, signal-to-noise ratio(SNR) and contrast-to-noise ratio (CNR) were obtained. Influence of different tube voltages and simultaneous iterative reconstruction levels on image quality was investigated. Results The CTDIvol, DLP and E were 12.50 mGy, 453 mGy ·cm, and 6.35 mSv, at 120 kV; 7.60 mGy, 276.50 mGy ·cm and 3.87 mSv at 100 kV; 3.80 mGy, 138.70 mGy ·cm and 1.94 mSv at 80 kV, respectively. Radiation dose significantly decreased as the tube voltage decreased. The tube voltage had significant influence on the CT values and its standard deviation(SD), SNR and CNR(F=1 029.24,8.56,3.44,4.09,P<0.05).The reconstruction algorithms had great influence on the SD, SNR and CNR as well(F=33.99,54.80,54.72,P<0.05), while no significant influence on CT values (P>0.05). There were no significant differences between the influences of tube voltages and reconstruction algorithms on the CT values of mediastinal soft tissues (P>0.05).Conclusions Using a protocol of 80 kV with a standard IR algorithm could result in reduction of the effective dose and high-quality thoracic CT angiogram.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11566205  On-line:0

v
Scan QR Code
&et=22F1D5E3CC41362B2C383DF5E4079622F117D5B43D1CC58AFE2D39152525579B783F0C85102577116167CA6DCB80ABAB840AF2A3298FA45F878425C7378A9EB070DAC9D450D676B6642435E9077E4786790624AF7F9A63667F3855F045E1468C8D9609381D0B9FBED66FE5DB4A4320EF4BB435ABF7486544AD4DEBBFF97893172AB5DA6F1A058C3B5F23F3C0EB7A909C996BC8B4F5BA14C3077589BE2E7760CA32D6C904369DB88CE9741BE6190A6BB2758915D6CF03B2F2&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=6C6113E1FCA1843691244D2131B1F29B&vid=&iid=94C357A881DFC066&sid=23F20F9780C3579E&eid=7D6CD8918B045FD4&fileno=20150520&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="6C6113E1FCA1843691244D2131B1F29B";