Yang Jinlei,Bu Mingwei,Huang Wei,Liu Boyu.Effects of horizontal beam on IMRT plan for lung cancer[J].Chinese Journal of Radiological Medicine and Protection,2015,35(5):353-355,359
Effects of horizontal beam on IMRT plan for lung cancer
Received:August 23, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.05.008
KeyWords:Horizontal beam  Intensity modulated radiation therapy  Lung cancer
FundProject:
Author NameAffiliationE-mail
Yang Jinlei 130012 长春, 吉林省肿瘤医院放疗研究室  
Bu Mingwei 130012 长春, 吉林省肿瘤医院放疗研究室  
Huang Wei 130012 长春, 吉林省肿瘤医院放疗二科  
Liu Boyu 130012 长春, 吉林省肿瘤医院放疗研究室 1525191900@qq.com 
Hits: 2383
Download times: 2256
Abstract::
      Objective To explore the effect of horizontal beam on IMRT plan for lung cancer. Methods Totally 18 patients were enrolled, including 10 cases of left lung cancer and 8 cases of right lung cancer. Two intensity modulated radiation therapy (IMRT) plans were generated: one was the normal IMRT plan and another was the IMRT plan including a horizontal beam (H-IMRT). MUs, sub-field and the dose distribution of target and organs at risk (OARs) were compared. Results There were no statistically significant differences between two plans in Dmax, Dmin and Dmean of PTV, the V30 and Dmean of the ipsilateral lung, V20 and V30 of the contralateral lung, and V30 of the total lung. When the horizontal beam was added, the PTV heterogeneity index(HI) and conformity index (CI) were significantly increased (t=-9.33, 10.88,P<0.05), the other dose index in lung was inferior (t=-1.55--7.58, P<0.05), whle heart sparing were clearly better (t=1.84-3.99,P<0.05), the number of MUs and segments were less(t=12.57,3.19, P<0.05). Conclusions IMRT in the treatment of lung cancer can further improve the PTV coverage when the horizontal beam is added. However, the increasing lung volume of low dose irradiation is not conducive to lung protection.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9074489  On-line:0

v
Scan QR Code
&et=970261AF74811A5DA21CD21E7E6ADC23D2A521E907366014036DB5F437135EEFFA1D391DA31D1C2457CE38EA1709574E1FD7E3C6ECB70E93&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=6C6113E1FCA18436496AB30F1505EC77&vid=&iid=94C357A881DFC066&sid=A5111BA190517959&eid=0464D5A41B4D9EE0&fileno=20150508&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="6C6113E1FCA18436496AB30F1505EC77";