Zhai Hezheng,Cheng Jinsheng,Ding Yanqiu,Guo Zhaohui,Qi Weihua,Wang Junliang,Liu Liming,Liu Ya,Zhang Honghong.Test of dosimetric characteristics of 2-D matrix detector in photon beams[J].Chinese Journal of Radiological Medicine and Protection,2015,35(2):144-148
Test of dosimetric characteristics of 2-D matrix detector in photon beams
Received:May 23, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.02.017
KeyWords:2-D matrix detector  Dosimetric characteristics  Quality control  IMRT
FundProject:卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Zhai Hezheng Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Cheng Jinsheng Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China chengjs3393@163.com 
Ding Yanqiu Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Guo Zhaohui Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Qi Weihua 军事医学科学院附属医院  
Wang Junliang 军事医学科学院附属医院  
Liu Liming Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Liu Ya Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhang Honghong Key Laborarory of Radiological Protection and Nudear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Hits: 3395
Download times: 2515
Abstract::
      Objective To explore the dosimetric characteristics of 2-D matrix detector, and to establish method of quality control test.Methods By reference to IEC 60731-2011 requirements on ionization chamber dosimetric indicators, a range of dosimetric indicators of 2-D matrix detector were tested by using photon beams (60Co γ-rays, 6,10 and 15 MV X-rays),including dosimetric repeatability, dose response linearity, dose rate dependency and dose measuring uniformity.Results The reproducibility of 2-D detector was less than 0.3%, and the medium-and long-term stability was less than 0.5%. The dose-response relationship was linear in the range of 10-250 cGy for 6,10,15 MV beams and 60Co photons. The dose linearity was less than 0.5%. The maximum deviation in measurement values was 0.87% in the dose rate range of 200 to 600 MU/min. The dose response derivations between different ionization chambers were less than 1.0%.Conclusions 2-D matrix detector has good dosimetric characteristics and the quality control testing method is feasible.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9078580  On-line:0

v
Scan QR Code
&et=74341BBFBC49985B5663F95F199AB18A1F474272CFF2AB7FB032523FFF6172A946C288713DA71D1BA4F2434D4B122FDF8EFAF39286124F752B2DE523CE9D1C7E6342B390584B8D2404CEBDE9CA792077&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=8A8C7B437C22286A61E729789AF51591&vid=&iid=0B39A22176CE99FB&sid=3986B25773CB6C30&eid=856C2E13D1000DB7&fileno=20150217&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="8A8C7B437C22286A61E729789AF51591";