.Investigation and analysis of patient dose levels from diagnostic radiology in Beijing[J].Chinese Journal of Radiological Medicine and Protection,2014,34(9):692-695
Investigation and analysis of patient dose levels from diagnostic radiology in Beijing
Received:December 10, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.09.013
KeyWords:Diagnostic radiology  Radiotherapy  Radiation dose  Quality control (QC)
FundProject:卫生行业科研专项(201002009)
Author NameAffiliationE-mail
娄云 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
王宏芳 Beijing Center for Disease Control and Prevention, Beijing 100013, China wanghf@jlu.edu.cn 
万玲 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
冯泽臣 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
马永忠 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
张泓 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
王新明 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
朱维杰 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
武大鹏 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
韩浚 Beijing Center for Disease Control and Prevention, Beijing 100013, China  
Hits: 4037
Download times: 2482
Abstract::
      Objective To investigate the radiation dose levels to the adults examined from diagnostic exposure in Beijing.Methods The radiation doses to the examined individuals were measured by using individual diagnostic radiology equipments in 30 random hospitals from a total of 10 districts and suburban areas, including 1 182 samples of X-ray photography,542 samples of mammography and 410 samples of CT examination.Results 2 134 samples were measured in this study. The dose ranges of X-ray photography, CR, and DR were 0.4-24.1, 0.3-13.9 and 0.1-15.9 mGy, respectively. The average dose range of glandular breast was 0.3-5.4 mGy. In 410 CT samples the value of CTDIw, CTDIvol and DLP were 28.1-96.3 mGy, 7.0-23.4 mGy, and 162.2-898.1 mGy ·cm, respectively. Conclusions Several dose levels from diagnostic examination were higher than guidance level for medical exposure in GB 18871-2002,which should be noted.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9076120  On-line:0

v
Scan QR Code
&et=9A2B3ECE8B55D590AE99BF1AE9328FD22387E875AEADCACCB728CA6D687B6943DC6FEEF6746E337D91D46EAF223FB19B70DDAC6C3DAA8C86F672FA4AC8F9B7CC05497B90CB44C8703DECD376341D7AD0EDCC8C98A513BA55&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=489B9E8FA8E68A9D4995D532DDD929CE&vid=&iid=9CF7A0430CBB2DFD&sid=15F713CDE16A589C&eid=08B2E838F29A693A&fileno=20140913&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="489B9E8FA8E68A9D4995D532DDD929CE";