Lu Feng,Chen Rui,Shang Yunying,et al.Flushing-free film test of 192Ir accuracy of position and step distance for afterloading systems[J].Chinese Journal of Radiological Medicine and Protection,2014,34(7):543-546
Flushing-free film test of 192Ir accuracy of position and step distance for afterloading systems
Received:September 12, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.07.017
KeyWords:Flushing-free film  Afterloading systems  192Ir source  Position accuracy  Step distance accuracy
FundProject:山东省科技发展计划(2009GG10002051)
Author NameAffiliationE-mail
Lu Feng Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China  
Chen Rui Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China  
Shang Yunying Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China  
Chen Yue Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China  
Min Nan Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China  
Chen Yingmin Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China  
Deng Daping Institute of Radiation Medicine, Shandong Academy of Medicine Sciences, Jinan 250062, China yanzhid@126.com 
Hits: 4164
Download times: 2509
Abstract::
      Objective To study the method of measuring the position accuracy and the step distance accuracy of afterloading system with 192Ir source by using flushing-free film.Methods The position accuracy and the step distance accuracy of a China-made afterloading system with 192Ir source was measured by using GAFCHROMIC® EBT3 flushing-free film. The film was scanned to proper image format, required by dose analysis software, by EPSON PREFACTION V700 PHOTO scanner. Then images are analyzed by using film dose analysis software in SNC Patient 5.2. Results With focus on the center of active section of source, the position accuracy of this afterloading system with 192Ir source was -0.75 mm. Using film analysis could make the step point to tell apart if the step distance was 5 mm away by the method of film analysis, but couldnot make it to tell apart if the step distance was 2.5 mm away. The 2.5 mm step distance accuracy could be judged if the distance between the 1st point and the 3rd point was 5 mm, then the 2.5 mm step distance could be deemed to no deviation. The 5 mm step distance of this afterloading system had no deviation in continuous 9 step points measured by flushing-free film. The indirect measuring results of the 2.5 mm step distance had no deviation as well. The position accuracy of this afterloading system measured with the flushing-free film accorded with the national standards. Conclusions The method of measuring the position accuracy and the step distance accuracy of the afterloading system with 192Ir source by using flushing-free film is technically feasible.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11565688  On-line:0

v
Scan QR Code
&et=9E340EB12EFCA825F137B2F36D9D21EAB6845EF39B5F6B5A6990282F18DF3B8A2FF52CA4995318AE264C7A0AB913F2B3035B849103D0A8614E087AD81C8AA8CAAD1470BC9A7292AFCE4E3E145F21860FBF8EFE6435668DFD3D3FBAF205E7E8A80965E893422F033ED107F67708DCA88C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=0430BD2A3227970BD6E1BE7A369F007F&vid=&iid=DF92D298D3FF1E6E&sid=BA305A52E2EE9350&eid=DEBDB7F30FBA7F9B&fileno=20140717&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="0430BD2A3227970BD6E1BE7A369F007F";