Ma Xinxing,Luo Suming,He Zhijian,Zhou Wenshan.Verification of dosimetric methodology for auditing radiotherapy quality under non-reference condition in Hubei province[J].Chinese Journal of Radiological Medicine and Protection,2014,34(6):461-465
Verification of dosimetric methodology for auditing radiotherapy quality under non-reference condition in Hubei province
Received:December 20, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.06.016
KeyWords:Non-reference condition  Finger ionization chamber  Photon beams  Electron beams  Absorbed dose
FundProject:IAEA基金资助项目(11916/R0);卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Ma Xinxing Hubei Provincial Center for Disease Control and Prevention, Wuhan 430079, China  
Luo Suming 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室  
He Zhijian 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室  
Zhou Wenshan Hubei Provincial Center for Disease Control and Prevention, Wuhan 430079, China 14212423@qq.com 
Hits: 3518
Download times: 2200
Abstract::
      Objective To verify the reliability of TLD-based quality audit for radiotherapy dosimetry of medical electron accelerator in non-reference condition by monitoring the dose variations from electron beams with different field sizes and 45° wedge and the dose variations from photon beams with different field sizes and source-skin distance. Methods Both TLDs and finger ionization chambers were placed at a depth of 10 cm in water to measure the absorbed dose from photon beams, and also placed at the depth of maximum dose from electron beams under non-reference condition. TLDs were then mailed to National Institute for Radiological Protection, China CDC for further measurement. Results Among the 70 measuring points for photon beams, 58 points showed the results with a relative error less than ±7.0%(IAEA's acceptable deviation: ±7.0%)between TLDs and finger ionization chambers measurements, and the percentage of qualified point numbers was 82.8%. After corrected by Ps value, 62 points were qualified and the percentage was up to 88.6%. All of the measuring points for electron beams, with the total number of 24, presented a relative error within ±5.0% (IAEA's acceptable deviation: ±5.0%) between TLDs and finger ioization cylindrical chambers measurements. Conclusions TLD-based quality audit is convenient for determining radiotherapy dosimetric parameters of electron beams in non-reference condition and can improve the accuracy of the measuring parameters in connection with finger chambers. For electron beams of 5 MeV<E0<10 MeV, the absorbed dose parameters measured by finger ionization chambers, combined with TLD audit, can help obtain the precise and reliable results.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061651  On-line:0

v
Scan QR Code
&et=BF7E82E5F022C752D74314D26FEFC0EA371A25A9E8DBA0230D0BE56B6773964318C59A8A90362486B1F5702C3DAB8B78B4318F94747B1FA6F3E7F065A3294ABD438D87119388BAAFBBFE3FEE6DB998EC28694DA3182A481203341126AC9C30F703295F8C4B52C3998DC68D747549C79D791405DA9F6C3698DFE9E86C31D1721C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=BBCD330085D72EC4E681BF3E95E1BD4D&vid=&iid=B31275AF3241DB2D&sid=FCACFF68346F8D4F&eid=BC88D6B0750E09D1&fileno=20140616&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="BBCD330085D72EC4E681BF3E95E1BD4D";