Cui Yong,Liu Baochen,Chen Zhongxing,Guo Junqiao,Luo Suming,He Zhijian,Yuan Jilong.Verification of dose parameters under reference and non-reference conditions for radiotherapy in Liaoning province[J].Chinese Journal of Radiological Medicine and Protection,2014,34(6):456-460
Verification of dose parameters under reference and non-reference conditions for radiotherapy in Liaoning province
Received:November 25, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.06.015
KeyWords:TLD  Reference and non-reference conditions  Finger ionization chamber  Plane parallel ionization chambers  Photon beams  Electron beams
FundProject:IAEA基金资助项目(11916/R0);卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Cui Yong Liaoning Center for Disease Control and Proevention, Shenyang 110005, China  
Liu Baochen Liaoning Center for Disease Control and Proevention, Shenyang 110005, China  
Chen Zhongxing Liaoning Center for Disease Control and Proevention, Shenyang 110005, China  
Guo Junqiao Liaoning Center for Disease Control and Proevention, Shenyang 110005, China guojunqiao@lncdc.com 
Luo Suming 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室  
He Zhijian 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室  
Yuan Jilong 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室  
Hits: 3052
Download times: 2187
Abstract::
      Objective To verify the reliability of dose parameters of radiotherapy under reference and non-reference conditions by using TLD. Methods Dose parameters were verified by using TLDs under reference and non-reference conditions, including the maximum dose in axel of 5 electron beams with energy of 9 MeV and the variations of dose by depth, source-skin distance, exposure field and 45° wedge for 10 photon beams with energy of 6 MV in 5 hospitals. Results The average relative deviation of 6 MV photon beam measured between TLDs and finger ionization chambers were 4.45%, within ±7% as required by IAEA. The average relative deviation of 9 MeV electron beam measured between TLDs and plane parallel chambers were 2.45%, within ±5% was required by IAEA. Conclusions Measuring dosimetric parameters by using TLDs under reference and non-reference conditions was reliable and feasible.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9058575  On-line:0

v
Scan QR Code
&et=BF7E82E5F022C752D74314D26FEFC0EA5F49472FD56768E2E931E6F65EEBADF8F484241B9E32B65503A51FEF742141DC3F8B43024E3674426F24A168C6561D2DB4ADB1B1ECE499516E0568B62B68550D9AD7F9DA5858A3FE58691AE3E50BAB746395B13575AD91F16A565E13859D326FB91BC9999784181C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=E9D710F0D9B2C145E61E5639698D1EEB&vid=&iid=B31275AF3241DB2D&sid=522844664D9E629A&eid=E0172F1A638CE984&fileno=20140615&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="E9D710F0D9B2C145E61E5639698D1EEB";