Cheng Xiaojun,Hu Chuanpeng,Qiao Hongbing,Wei Kunjie,Luo Suming,He Zhijian.TLD-based quality audit for radiotherapy dosimetry under reference and non-reference condition in Henan province[J].Chinese Journal of Radiological Medicine and Protection,2014,34(6):446-450
TLD-based quality audit for radiotherapy dosimetry under reference and non-reference condition in Henan province
Received:November 07, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.06.013
KeyWords:TLD  Reference and non-reference conditions  Finger ionization chamber  Plane parallel ionization chamber  Photon beam  Electron beam
FundProject:IAEA基金资助项目(11916/R0);卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Cheng Xiaojun Henan Institute of Occupational Medicine, Zhengzhou 450052, China  
Hu Chuanpeng Henan Institute of Occupational Medicine, Zhengzhou 450052, China  
Qiao Hongbing Henan Institute of Occupational Medicine, Zhengzhou 450052, China  
Wei Kunjie Henan Institute of Occupational Medicine, Zhengzhou 450052, China  
Luo Suming 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室 sumingluo@163.com 
He Zhijian 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室  
Hits: 3736
Download times: 2210
Abstract::
      Objective To verify the reliability of radiotherapy dosimetric parameters in reference and non-reference conditions using thermoluminescent dosimeters (TLDs). Methods Using the established TLD method, the dose variations with different radiation field sizes and 45° wedge plate were verified for 10 photon beams of 6 MV, together with dosimetric parameters at the point of maximum axial dose for 4 electron beams of 9 MeV under reference and non-reference conditions. Comparisons were made between TLD results and finger ionization chamber results. Results The average relative deviation, for 6 MV photon beams, between TLD results and finger ionization chamber measurements was 4.7%, within±7% as required by the IAEA. The average relative deviation, for 9 MeV electron beam, between TLD results and plane parallel ionization chamber measurements was 2.4%, not beyond ±5% permitted by IAEA. Conclusions Using TLD method to verify the radiotherapy dosimetric parameters in reference and non-reference conditions was reliable, simple and feasible.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9064097  On-line:0

v
Scan QR Code
&et=20416C467120962B85B24BA180C1AF8F9E5E894F9724F0342F2E6A408642DB6A158F34013B42E3E0D0F89DD172F10E433C9F34EBFC586F29FB5B368D49DCF231602AF3D78C82141B2E914D743530E4451F1C430F016121CD724A3B7FAEE5977EF37FF9A4CDA56EA1FC174F6DF1152E372C2DA23492D71226&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=E9D710F0D9B2C1459B8502876FD979A5&vid=&iid=B31275AF3241DB2D&sid=BD7D27247C63490C&eid=0A0BD3F594F876C0&fileno=20140613&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="E9D710F0D9B2C1459B8502876FD979A5";