Luo Suming,He Zhijian,Li Kaibao,Xue xian.Development of methodology for auditing dosimetric parameters for photon and electron beams under non-reference conditions[J].Chinese Journal of Radiological Medicine and Protection,2014,34(6):441-445
Development of methodology for auditing dosimetric parameters for photon and electron beams under non-reference conditions
Received:December 26, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.06.012
KeyWords:Non-reference condition  TLD  Photon and electron beams  Plane parallel chamber  Absorbed dose
FundProject:IAEA基金资助项目(11916/R0);卫生行业科研专项(201002009)
Author NameAffiliation
Luo Suming Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China 
He Zhijian Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China 
Li Kaibao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China 
Xue xian Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China 
Hits: 3790
Download times: 2112
Abstract::
      Objective To develop a methodology for auditing dosimetric parameters using TLD for photon and electron beams in non-reference conditions. Methods The correction experiments have been developed for non-linear dose response, PAAM holder, energy, fading and dispersion using TLD for 60Co γ-ray, high energy X-ray and electron beams.The measuring method was set up for absorbed dose estimation by TLDs in water. 3 photon beams and 2 electron beams were selected for research purposes, and 5 photon beams and 4 electron beams for reliability research. Results The research results were evaluated for 60Co source,6, 10, 15 and 18 MV photon beams in non-reference conditions at off-axis, with the relative deviation within -0.1%-7.2% at-off axis (IAEA's acceptable deviation: ±7.0%). The verification results were evaluated for 5 photon beams in non-reference conditions at on-axis, with the relative deviation are within 0.1%-7.0%. The verification results were evaluated for 4 electron beams in non-reference conditions, with the relative deviation within 0-4.7% (IAEA's acceptable deviation: ±5.0%). Conclusions It is convenient and accurate to use TLD method for quality audits for clinic dosimetry parameters in radiotherapy. Two kinds of IAEA TLD holders are feasible for use in TLD audits. Absorbed doses for high energy electron beam were corrected using plane parallel chambers and verified using TLD, with good results obtained.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9066647  On-line:0

v
Scan QR Code
&et=A90EAC7419FB9B22C5C57F2085C69769189A7B80470D296C2D8873CF57BA7F5CE819769CDD07C150521FECED4BEFBC41BDFFEAF33135C686D6A604E2A37DCF5CDFDA343CDA3FA2D2C876433DEEE75FE4185A6FA14499F66DDA4B38947B02A9388A20F01AF92D8FD0194207B5DBC78E037C0D6AF692CA612E22ABCCDB6F889D49&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=E9D710F0D9B2C145C1A642FBEFFCB1A1&vid=&iid=B31275AF3241DB2D&sid=5319469C819FCFF1&eid=A48DE16C07AAAB06&fileno=20140612&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="E9D710F0D9B2C145C1A642FBEFFCB1A1";