Xing Xiaofen,Chu Xuegang,Cui Tong,Zheng Xuliang,Guo Ruisong.A clinical test and application research of IMRT 3D dose verification system[J].Chinese Journal of Radiological Medicine and Protection,2014,34(6):427-430
A clinical test and application research of IMRT 3D dose verification system
Received:July 11, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.06.007
KeyWords:Intensity-modulated radiotherapy  Quality assure  Three-dimensional dose verification  Dosimetry
FundProject:
Author NameAffiliationE-mail
Xing Xiaofen Department of Radiation Oncology, Shanxi Tumor Hospital, Taiyuan 030013, China  
Chu Xuegang Department of Radiation Oncology, Shanxi Tumor Hospital, Taiyuan 030013, China  
Cui Tong Department of Radiation Oncology, Shanxi Tumor Hospital, Taiyuan 030013, China  
Zheng Xuliang Department of Radiation Oncology, Shanxi Tumor Hospital, Taiyuan 030013, China  
Guo Ruisong Department of Radiation Oncology, Shanxi Tumor Hospital, Taiyuan 030013, China sxgrs@163.com 
Hits: 3744
Download times: 2278
Abstract::
      Objective To test the accuracy of a three-dimensional dose verification system CompassR, which reconstructing dose distribution based on measurements and independent dose calculation, and to evaluate the feasibility of its application in clinical intensity-modulated radiotherapy(IMRT) quality assure. Methods A set of square-wave chart patterns of 2 cm, 1 cm and 0.5 cm gaps was designed and 11 completed IMRT lung plans were selected for the test. EDR2 film and the ionization chamber were used for test and verifying of plane dose distribution and some special points dose of CompassR. The IMRT phantom plans were verified by CompassR with three-dimension based on anatomical information. Parameters including the volume γ pass rate and the average dose deviation were tested using dose volume histograms. Results In square-wave chart patterns test, the dose distribution reconstructed and calculated by CompassR coincided with the measurement using film. The γ pass rates (3%/3 mm,2%/2 mm) exceeded 90%. When the width of field is 0.5 cm, the γ pass rate was a little lower on account of the penumbra zone. Compared to the dose distribution profile which was measured by film, the maximum deviations of the dose distribution profile which was reconstructed and calculated by CompassR were 3.21% and 2.70%. The absolute dose deviation of specific point in the IMRT plans was less than 3%,the maximum deviation occurred in the lung. Compared to film, the averageγpass rates on the isocenter plain in IMRT plan were (94.65+1.93)% (3%/3 mm) which was reconstructed by CompassR. In three-dimensional dose verification, the volume γ pass rates of targets and risk organs were not less than 90%, and the deviation of average dose was less than 1%. Conclusions Accuracy of the tested system satisfies the demand of IMRT dose verification. CompassR could provide information of volumetric dosimetry and anatomical location of dose error, which is benefit for evaluating the clinical value of verification result.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9060391  On-line:0

v
Scan QR Code
&et=232A82BFAD75C48376EF36834C5A66F59EDD59D3E1032B9B58CB89DE146E70922882D72E1456DC5FE446D9BE3B12CB9895EEEFE129D42BD5B45630A6054D5CA2D89F61500FAF25F80A2F8D13FAAD4E8A&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=E9D710F0D9B2C145AA19E53D77CEEA5E&vid=&iid=B31275AF3241DB2D&sid=CB3428B1EFB1C133&eid=389DA78D878702A9&fileno=20140607&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="E9D710F0D9B2C145AA19E53D77CEEA5E";