Cai Jinfeng,Lin Hui,Cao Ruifen,Dai Yumei,Pei Xi,Zhang Jun,Xu Gongming.Realization of dose verification tool for IMRT plan based on DPM[J].Chinese Journal of Radiological Medicine and Protection,2014,34(5):381-385
Realization of dose verification tool for IMRT plan based on DPM
Received:November 04, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.05.017
KeyWords:Monte Carlo method  Intensity modulated radiotherapy  Dosimetry  Dose verification
FundProject:中国科学院战略性先导科技专项(XDA03040000);国家自然科学基金 (81101132)
Author NameAffiliationE-mail
Cai Jinfeng School of Electronic Science & Application Physics, Hefei University of Technology, Hefei 230009, China  
Lin Hui School of Electronic Science & Application Physics, Hefei University of Technology, Hefei 230009, China huilin@hfut.edu.cn 
Cao Ruifen 中国科学院核能安全技术研究所  
Dai Yumei School of Electronic Science & Application Physics, Hefei University of Technology, Hefei 230009, China  
Pei Xi 中国科学院核能安全技术研究所  
Zhang Jun 中国科学院核能安全技术研究所  
Xu Gongming 中国科学院核能安全技术研究所  
Hits: 3088
Download times: 2122
Abstract::
      Objective To build a Monte Carlo dose verification tool for IMRT Plan by implanting an irradiation source model into DPM code and to extend the ability of DPM to calculate any incident angles and irregular-inhomogeneous fields.Methods The virtual source and the energy spectrum unfolded from the accelerator measurement data were used, in combination with optimized intensity maps, to calculate the dose distribution of the irradiation irregular-inhomogeneous field.The irradiation source model of accelerator was substituted by a grid-based surface source.The contour and the intensity distribution of the surface source were optimized by IMRT.The dose calculation was realized by combining the position of the emitter with the fluence map from the IMRT plan.The weight of the emitter was decided by the grid intensity.The direction of the emitter was decided by the combination of the virtual source and the emitting position.The weighted fraction of the emitter was also combined with the flux grid intensity based on the particle transport model of DPM code.Results The accuracy of calculation was verified by comparing with the measured data.It was illustrated that the differences were acceptable (<2% inside the field, 2-3 mm in the penumbra).The dose calculation of irregular field by DPM simulation was also compared with that of FSPB (Finite Size Pencil Beam).The passing rate of gamma analysis was 95.1% for peripheral lung cancer.The regular field and the irregular rotational field were all within permissible range of error.The calculation time of regular fields were less than 2 h, and that of the test of peripheral lung cancer was 160 min.Conclusions The adapted DPM code with its simple irradiation source model is faster than that with classical Monte Carlo procedure.Its computational accuracy and speed satisfy the clinical requirement, and it can be useful as a Monte Carlo dose verification tool for IMRT Plan.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061024  On-line:0

v
Scan QR Code
&et=066A139AF21709FBB133F9DDA864152ED35FC8BCA7F4BA4C846697EE56980D7F6415174FE5E48453CE1F3ADF65D2AAF2FE9C3D25B04241C13FF1B8B44712B97383265CFDE82E5FCB&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=3CC66326E61FC91B761B45591350D8DA&vid=&iid=94C357A881DFC066&sid=B9B90065CF5CD7F0&eid=44FDB9366EDDFA2B&fileno=20140517&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="3CC66326E61FC91B761B45591350D8DA";