Qiu Zhengshuai,Deng Daping,Li Quantai,et al.Investigation on radiation doses to patients in digital radiography[J].Chinese Journal of Radiological Medicine and Protection,2014,34(5):372-375
Investigation on radiation doses to patients in digital radiography
Received:October 26, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.05.015
KeyWords:Digital radiography  Exposed patients  Radiation dose
FundProject:卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Qiu Zhengshuai School of Medicine and Life Sciences, University of Jinan-Shangdong Academy of Medical Sciences, Jinan 250062, China  
Deng Daping 山东省医学科学院放射医学研究所  
Li Quantai 山东省医学科学院放射医学研究所  
Song Gang 山东省医学科学院放射医学研究所  
Su Xu 中国疾病预防控制中心辐射防护与核安全医学所 辐射防护与核应急中国疾病预防控制中心重点实验室 suxu@nirp.cn 
Hits: 5749
Download times: 2445
Abstract::
      Objective To investigate the patients' radiation dose received in digital radiography(DR) and provide basic data for developing diagnostic reference levels.Methods The patient's ESD was estimated using the TLDs and DAP was measured by the dose-area product meter.The E values were then calculated by the DAP using Monte Carlo data and RefDose software.Measurements were made for twelve types of examination: skull PA, skull LAT, chest PA, chest LAT, abdomen AP, pelvis AP, cervix spine PA, cervix spine LAT, thoracic spine PA, thoracic spine LAT, lumber spine PA and lumber spine LAT.Results Both kV and mAs varied in the same type of examination for ESD, DAP and E(F=33.47, 24.68, 43.19,P<0.05).The dose each time for lumber spine LAT was the highest, reached 4.62 mGy in ESD and 2.26 Gy·cm2 in DAP, respectively.The E of abdomen AP averaged as 0.59 mSv, higher than that of lumber spine LAT.Even for the same type of examination, the dose from each equipment was different.Conclusions DR has the potential to reduce the patients' radiation doses.The guidance levels suitable for Chinese population should be established as soon as possible.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11565368  On-line:0

v
Scan QR Code
&et=9A2B3ECE8B55D590F309F17E6178CC7F8A4A1EAB25E97F6330F38B817798CF46B411E8787696039326A75911B6AB58183EF7BF23BF68643A9DAD480F94471BD1A915E12C7F0894AF&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=3CC66326E61FC91B4F152CBFC4984909&vid=&iid=94C357A881DFC066&sid=965F4E89CD0AFC30&eid=315A3D008C6ECFC8&fileno=20140515&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="3CC66326E61FC91B4F152CBFC4984909";