Ying Yuanning,Zhang Liren,Xu Dongsheng,Liu Zhe,Liu Junbo,Guan Yanfang,Fu Donghai,Zhou Wei,Liu Yanping.Clinical research of dose reduction protocol in coronary artery CT angiography with discovery 750 HD[J].Chinese Journal of Radiological Medicine and Protection,2014,34(4):310-313
Clinical research of dose reduction protocol in coronary artery CT angiography with discovery 750 HD
Received:May 22, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.04.019
KeyWords:Dose reduction  Coronary angiography  Tomography, X-ray computed
FundProject:
Author NameAffiliationE-mail
Ying Yuanning Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Zhang Liren Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China cjr.zhangliren@vip.163.com 
Xu Dongsheng Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Liu Zhe Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Liu Junbo Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Guan Yanfang Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Fu Donghai Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Zhou Wei Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Liu Yanping Department of Radiology, TEDA International Cardiovascular Hospital, Tianjin 300457, China  
Hits: 3060
Download times: 2676
Abstract::
      Objective To evaluate effects on image quality and effective radiation with the dose reduction protocol adapted by body mass index (BMI) in coronary artery CT angiography with discovery 750 HD. Methods Two hundred and seven patients suspected with coronary artery disease (CAD) were enrolled in Group A who were underwent CT angiography with routine radiation, and another 242 patients were enrolled in Group B who received a new scanning protocol with tube voltage and tube current modulation depending on the BMI. The scores of image quality, signal noise ratio(SNR), CT dose index (CTDIvol), and effective dose (E) were compared and the correlation between the scanning parameters and E was analyzed. Results The mean radiation dose in Group B(1.2, ranging from 0.8 to 2.5 mSv)was significantly lower than that in Group A(1.9,ranging from 1.5 to 2.3 mSv; Z=-6.24,P<0.05). The multivariate regression analysis revealed that the first factor affecting effective radiation dose ranked as scanning mode, tube voltage, tube current, and scan length, respectively. The image quality between two groups was not significantly different. The SNR in Group B was greater than that in Group A (Z=-2.22,P<0.05).Conclusions Dose reduction protocol in coronary artery CT angiography with discovery 750 HD would help to reduce the radiation dose effectively without sacrificing image quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9062075  On-line:0

v
Scan QR Code
&et=3C06DB566315F8DCD345A9AB065A4DDCEE0F785E10A08824C1AE09352CD2DDCC07F55C723A5AD5228F570DB39A55A988C8740B41D164F967BEFA71588D57006A228F6FF64E3E3E7EA963329E7DB3FE9C42900B806D0874AAF5B1B03929F7F97CF18E562273C127CD&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=6254F803FD8F6343082AF35B92F98E82&vid=&iid=E158A972A605785F&sid=85002451B65CE0D1&eid=C2F76551C0111538&fileno=20140419&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="6254F803FD8F6343082AF35B92F98E82";