Xu Hui,Xue Xian,Zhao Hongfeng,Wang Jianchao,Wei Kedao,Yue Baorong.Peak skin dose measurements in two cardiac interventional procedures using radiochromic film[J].Chinese Journal of Radiological Medicine and Protection,2014,34(4):302-305
Peak skin dose measurements in two cardiac interventional procedures using radiochromic film
Received:June 10, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.04.017
KeyWords:Cardiac interventional  Cumulative dose  Peak skin dose  Deterministic effects
FundProject:卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Xu Hui National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Xue Xian National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhao Hongfeng 北京市朝阳区疾病预防控制中心  
Wang Jianchao National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Wei Kedao National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Yue Baorong National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China bryue@163.com 
Hits: 3383
Download times: 2626
Abstract::
      Objective To measure the peak skin dose (PSD) in two cardiovascular interventional procedures, including coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) using radiochromic film. Methods Gafchromic XR-RV3 film was selected to measure PSD in two hospitals. The films were placed on the table underneath the patient during interventional surgery. The kV, mA, fluoroscopy time, dose-area product (DAP), and cumulative dose at reference point and other relevant information were recorded for all cases. Using the Epson V750 flatbed scanner for scanning and analyzing film,FilmQA software was chosen to analyze the pixel value of red, green and blue color channels. The PSD was determined using red channel data. The correlation and linear regression analysis between PSD and device-displayed parameters was carried out. Results PSD were measured using XR-RV3 film for 26 CA and 19 CA+PTCA procedures. For CA procedures, maximum fluoroscopy time, cumulative dose and DAP were 17.62 min, 1 498.50 mGy and 109.68 Gy ·cm2, respectively. The maximum PSD was 361.20 mGy. However, for CA+PTCA procedures, maximum fluoroscopy time, cumulative dose and DAP were 64.48 min, 6 976.20 mGy and 5 336.00 Gy ·cm2, respectively. One patient with CA+PTCA procedures was found to have received the PSD value more than 2 Gy, up to 2 195.70 mGy. DAP was found to be a good indicator (R2=0.815,P<0.05) of PSD for CA procedure, and correlated with cumulative dose (R2=0.916,P<0.05) for CA+PTCA procedures. Conclusions The PSD value of some patients in cardiac interventional procedures would exceed 2 Gy, the threshold of deterministic effects recommended by ICRP. The dose-related parameters value showed on DSA device can only used to estimate PSD roughly. Using XR-RV3 film accurate measurement of the PSD in interventional projects is a very fast and effective method.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9060890  On-line:0

v
Scan QR Code
&et=A32799621C139CD9E927033EC1EDC6872451EE6042EA29C50C3582E3E5459B2195C161F9B288277816F68E63663EC49A5782FE0E6D074A8683340EB9985E4C6CBEB08D131A175ABF1E4F78C622610C4C93BF7C5B04351A4D374A84CCE58CEAF2&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=6254F803FD8F63435C2D834DB99FD1F4&vid=&iid=E158A972A605785F&sid=119B6C0AA09DE6B9&eid=407C905D8F0449C4&fileno=20140417&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="6254F803FD8F63435C2D834DB99FD1F4";