Zhang Jiyong,Lu Jiayang,Wu Lili,et al.Clinical application of IMSure QA software[J].Chinese Journal of Radiological Medicine and Protection,2014,34(4):292-294 |
Clinical application of IMSure QA software |
Received:June 26, 2013 |
DOI:10.3760/cma.j.issn.0254-5098.2014.04.014 |
KeyWords:Intensity modulated radiotherapy Dose verification IMSure QA MatriXX |
FundProject: |
Author Name | Affiliation | E-mail | Zhang Jiyong | Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China | | Lu Jiayang | Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China | | Wu Lili | Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China | | Chen Zhijian | Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China | zjchenmd@stu.edu.cn |
|
Hits: 3725 |
Download times: 1919 |
Abstract:: |
Objective To compare IMSure QA and MatriXX for intensity modulated radiotherapy plans, and to investigate the credibility of IMSure QA software.Methods Ten patients receiving intensity modulated radiotherapy were randomly chosen for the quality assurance plans with IMSure QA and MatriXX. Results The overall average of percentage pass points in 3% and 3 mm were (98.1±0.8)% with IMSure QA, and (97.9±0.6)% with MatriXX(t=0.86,P>0.05). Conclusion IMSure QA can be a reliable verification tool for IMRT QA. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|