Wen Fang,Hu Chunhong,Hu Su,Dai Qichun.The feasibility study of CT perfusion imaging of rabbit brain with 70 kV tube voltage[J].Chinese Journal of Radiological Medicine and Protection,2014,34(2):91-94
The feasibility study of CT perfusion imaging of rabbit brain with 70 kV tube voltage
Received:August 12, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2014.02.004
KeyWords:Low dose  Tomography  X-ray computed  Rabbit brain perfusion imaging  Effective dose
FundProject:国家自然科学基金(31271066,81171393);江苏省高教优势学科建设工程资助项目(PAPD-2011-0318)
Author NameAffiliationE-mail
Wen Fang Imaging Center, First Affiliated Hospital of Soochow University, Suzhou 215006, China  
Hu Chunhong Imaging Center, First Affiliated Hospital of Soochow University, Suzhou 215006, China hch5305@163.com 
Hu Su Imaging Center, First Affiliated Hospital of Soochow University, Suzhou 215006, China  
Dai Qichun Imaging Center, First Affiliated Hospital of Soochow University, Suzhou 215006, China  
Hits: 2771
Download times: 1981
Abstract::
      Objective To evaluate the feasibility of low dose CT perfusion imaging of brain using the tube voltage of 70 kV. Methods Twenty New Zealand white rabbits underwent cerebral CT perfusion imaging with tube voltage 70 and 80 kV separately, and the interval between two scans was 24 h. The effective dosage (E), cerebral perfusion parameters (CBF, CBV, MTT) and image quality indicator (SNR) of middle cerebral artery of each protocol were acquired and compared statistically. Results The effective dosage of 70 kV group was 1.91 mSv, 34.8% lower than that of 80 kV group with the dose of 2.93 mSv. There were no significant differences between the cerebral perfusion parameters acquired at different corresponding ROI. SNR of 70 kV group and 80 kV group were (23.15±5.98) and (21.23±9.18) with no significantly statistical difference. Conclusions The effective dosage of 70 kV cerebral CT perfusion imaging could be dramatically reduced with no significant influence on the results of perfusion parameters and image quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9079927  On-line:0

v
Scan QR Code
&et=876B4A42438D29B34F8569CFA947824DEC63262CF53FCA37A001BC3FFD6D32761DC952F62B8D50206BD39E2997F01F4B25C24809C804CC3D8ADFBFF1FB6D3D5162B87CBB771D836925739C2927A0F386364275495BDF8B7E&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9EAD63ADE6B277ED&aid=0999E3A37C445912D855777DB6600C81&vid=&iid=0B39A22176CE99FB&sid=C753EB8AC8F551B9&eid=BB0EA31DB1B01173&fileno=20140204&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9EAD63ADE6B277ED"; var my_aid="0999E3A37C445912D855777DB6600C81";