ZHANG Wei,MA Zhao,SHAO Peng,JIANG Wei,GUAN Yu-min.Comparison of dosimetry in radiotherapy for craniospinal irradiation[J].Chinese Journal of Radiological Medicine and Protection,2013,33(5):519-523
Comparison of dosimetry in radiotherapy for craniospinal irradiation
Received:January 23, 2013  
DOI:10.3760/cma.j.issn.0254-5098.2013.05.016
KeyWords:Craniospinal  3D-CRT  sIMRT  Dosimetry
FundProject:
Author NameAffiliation
ZHANG Wei Department of Radiation Oncology, Yantai Yuhuangding Hospital, Yantai 264000, China 
MA Zhao Department of Radiation Oncology, Yantai Yuhuangding Hospital, Yantai 264000, China 
SHAO Peng Department of Radiation Oncology, Yantai Yuhuangding Hospital, Yantai 264000, China 
JIANG Wei Department of Radiation Oncology, Yantai Yuhuangding Hospital, Yantai 264000, China 
GUAN Yu-min Department of Radiation Oncology, Yantai Yuhuangding Hospital, Yantai 264000, China 
Hits: 3065
Download times: 1928
Abstract::
      Objective To explore the dosimetry of simplified intensity modulated radiotherapy(sIMRT)in the spinal component of craniospinal irradiation (CSI) compared with conventional plan (3D-CRT). Methods Five previously treated patients were adopted to generate 3D-CRT, 3- and 5-field sIMRT plans. The prescribed dose was 36 Gy to the planning target volume (PTV) in 20 fractions. The dose distributions of target volume and normal tissues, and homogeneity index (HI) were ananlyzed using the dose volume histogram. The total monitor units (MUs) were also analyzed to compare the delivery time indirectly. Results For 3D-CRT plans, the region of the neighbouring fields only received 70% of the prescribed dose, and the maximum dose delivered in the normal tissues was increased to approximately 140% of the prescribed dose. The homogeneity index (HI) of 3- and 5-field sIMRT were 0.09±0.01 and 0.08±0.01, respectively, superior to that of 3D-CRT (0.18±0.02,t=7.80, 7.65,P<0.05). The values of V10 for the heart of 3- and 5-field sIMRT were (8.4±1.9)% and (8.4±2.0)%, respectively, lower than that of 3D-CRT( (36.0±6.0)%(t=13.3,13.0,P<0.05).V20 for the thyroid gland were (12.4±1.5)% and (12.4±1.6)%, respectively, lower than (69.4±5.7)% of 3D-CRT(t=26.3, 26.4,P<0.05).V20 for the larynx were (17.2±1.2)% and (17.9±1.5)%, respectively, lower than that of 3D-CRT[(89.4±7.0)%(t=25.5, 26.5,P<0.05)].V30 for the NT was (4.4±1.4)%,(4.9±1.9)%, lower than that of 3D-CRT[(31.9±6.1)%(t=8.5,10.1,P<0.05)].The averaged values of total MUs for 3- and 5-field sIMRT were 1100±177 and 1160±204, respectively, higher than that of 3D-CRT(640±151). Conclusions Compared with the 3D-CRT CSI technique, 3- and 5-field sIMRT have better dose distribution and can improve target dose uniformity and protect the organs at risk.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9057584  On-line:0

v
Scan QR Code
&et=8C10CEA789A144B8CD11BBC4BD3E5794655139D31031C46DFBB5A34BB596EAA9EFE7C434DE9576AB929EFD3FFB6340C6B8CF19AFEE952F3D1C713042AB2D4E881306F98EEA1C358D&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FF7AA908D58E97FA&aid=435D3C49FB32FBCF7305CDE885602D02&vid=&iid=94C357A881DFC066&sid=89389F643CA3F778&eid=C81D738643975BB0&fileno=20130516&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FF7AA908D58E97FA"; var my_aid="435D3C49FB32FBCF7305CDE885602D02";