Radiation shielding for medical linear accelerator room with useful beam towards its maze
Received:October 25, 2012  
DOI:10.3760/cma.j.issn.0254-5098.2013.04.027
KeyWords:Medical linear accelerator  Shielding  Radiation protection
FundProject:卫生行业科研专项(201002009)
Author NameAffiliationE-mail
LI Hai-liang Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China  
LIANG Hong-ri 威海市立医院  
ZHU Jian-guo Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China  
CHEN Ying-min Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China  
GAO Ze-yu Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China  
LIU Fang Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China  
LI Quan-tai Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China quantaili@163.com 
Hits: 2727
Download times: 2819
Abstract::
      Objective To study the calculational method for the radiotherapy facilities of the medical linear accelerator's useful beam towars its maze. Methods The shielding calculation was made under the relevant national standards for a radiotherapy treatment room and compared with the test results. Results The dose rates at the maze entrance as calculated and measured were 89 and 86 μSv/h inside the maze door, as well as 5.7 and 6.2 μSv/h outside the maze door, respectively. The calculated results were consistent with measured results. Conclusions By comparison of calculated results with measured results, the accuracy of the theoretical calculation method could be verified.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9070864  On-line:0

v
Scan QR Code
&et=15C444C2D45CD59A79A132A3D94A4F3DAD7E1583ED6528D6BE5AA596F3BE14CC99C0BBB83A6B072D22CD01F1DDB7D0806F036EB323D81B1FD0BE8A81E2246F69E18806902721262FDF836B8DAEE223BB6D839D1E04780718E9D1E8AC4D7C15B1&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FF7AA908D58E97FA&aid=CBE5FA4F89B07936144717F842F0EB7F&vid=&iid=E158A972A605785F&sid=CB3428B1EFB1C133&eid=389DA78D878702A9&fileno=20130427&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FF7AA908D58E97FA"; var my_aid="CBE5FA4F89B07936144717F842F0EB7F";