LI Hai-liang,LIANG Hong-ri,ZHU Jian-guo,et al.Radiation shielding for medical linear accelerator room with useful beam towards its maze[J].Chinese Journal of Radiological Medicine and Protection,2013,33(4):427-430 |
Radiation shielding for medical linear accelerator room with useful beam towards its maze |
Received:October 25, 2012 |
DOI:10.3760/cma.j.issn.0254-5098.2013.04.027 |
KeyWords:Medical linear accelerator Shielding Radiation protection |
FundProject:卫生行业科研专项(201002009) |
Author Name | Affiliation | E-mail | LI Hai-liang | Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China | | LIANG Hong-ri | 威海市立医院 | | ZHU Jian-guo | Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China | | CHEN Ying-min | Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China | | GAO Ze-yu | Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China | | LIU Fang | Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China | | LI Quan-tai | Institute of Radiation Medical, Shandong Academy of Medical Sciences, Jinan 250062, China | quantaili@163.com |
|
Hits: 3476 |
Download times: 3257 |
Abstract:: |
Objective To study the calculational method for the radiotherapy facilities of the medical linear accelerator's useful beam towars its maze. Methods The shielding calculation was made under the relevant national standards for a radiotherapy treatment room and compared with the test results. Results The dose rates at the maze entrance as calculated and measured were 89 and 86 μSv/h inside the maze door, as well as 5.7 and 6.2 μSv/h outside the maze door, respectively. The calculated results were consistent with measured results. Conclusions By comparison of calculated results with measured results, the accuracy of the theoretical calculation method could be verified. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|