LI Peng,GAO Lin-feng,ZHUO Wei-hai,CHEN Bo.A survey and analysis of dose level for adults undergoing posteroanterior chest X-rays photography in Shanghai[J].Chinese Journal of Radiological Medicine and Protection,2013,33(1):68-71
A survey and analysis of dose level for adults undergoing posteroanterior chest X-rays photography in Shanghai
Received:August 17, 2012  
DOI:10.3760/cma.j.issn.0254-5098.2013.01.019
KeyWords:Medical exposure  Chest X-ray photography  Posteroanterior  Entrance surface dose  Diagnostic reference level
FundProject:国家自然科学基金(11075040);上海市公共卫生优秀学科带头人培养计划(GWDTR201206)
Author NameAffiliationE-mail
LI Peng Institute of Radiation Medicine, Fudan University, Shanghai 200032, China  
GAO Lin-feng 上海市疾病预防控制中心  
ZHUO Wei-hai Institute of Radiation Medicine, Fudan University, Shanghai 200032, China whzhuo@fudan.edu.cn 
CHEN Bo Institute of Radiation Medicine, Fudan University, Shanghai 200032, China  
Hits: 3557
Download times: 2416
Abstract::
      Objective To survey the current dose level for adults undergoing posteroanterior (PA) chest X-ray photography in Shanghai, and to analyze the main factors affecting the dose level. Methods The thermoluminescence dosimeters (TLDs) were deployed to measure entrance surface dose (ESD) in the center of radiated fields. According to the numbers of three types of X-ray equipment and their distribution among different grade hospitals in Shanghai, the ESDs were surveyed for 1032 adults undergoing examination of PA chest X-ray photography from 2008 to 2010, and the relevant information was also recorded. The differences of the mean ESDs among different grade hospitals, equipment types and photographic parameters were statistically analyzed. Results The arithmetic mean of the ESDs for adults undergoing examination of PA chest X-ray photography in Shanghai was 0.23 mGy, and the 75th percentile value was 0.29 mGy, which were all below the current China diagnostic reference level (0.40 mGy). The averaged ESDs were significantly correlated with the grade of hospital, the equipment types and the photographic parameters. Conclusions Compared with the value in the last century, the ESD for adults undergoing PA chest X-ray photography in Shanghai obviously decreases, and there are the potentials for further reduction.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061456  On-line:0

v
Scan QR Code
&et=73559858ED1EA9C22E7B77AEBE54154987F06F7523EA704C63EEC21B4E8A1DE60BC21D678B5D0AD687C9AB456A08596AD7E8C6C07DF01D0B8C06454BE753E17D95D609B865FB66EE036304D55DE91099AD9E462B09BE518DEF5064B8A456C3D6183CA80F0994E9F805E8F043B50223DC&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FF7AA908D58E97FA&aid=08F5EC0AA861649165C01B29EAF4F26F&vid=&iid=CA4FD0336C81A37A&sid=68D88C2FCF9C3098&eid=4F2F18DD6F870C2C&fileno=20130119&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FF7AA908D58E97FA"; var my_aid="08F5EC0AA861649165C01B29EAF4F26F";