HAN Jun,LI Qin,CAO Ting,LIANG Zhi-wen,CHEN Mi.The dose rate response characterization of four kinds of dosimeters for linear accelerator[J].Chinese Journal of Radiological Medicine and Protection,2012,32(5):527-529
The dose rate response characterization of four kinds of dosimeters for linear accelerator
Received:March 26, 2012  
DOI:10.3760/cma.j.issn.0254-5098.2012.05.023
KeyWords:Dose rate  Dose verification  Dosimeter  Response characterization
FundProject:
Author NameAffiliationE-mail
HAN Jun Oncology Center, Union Hospital, Huazhong University of Science and Technology, Tongji Medical College, Wuhan 430023, China  
LI Qin Oncology Center, Union Hospital, Huazhong University of Science and Technology, Tongji Medical College, Wuhan 430023, China lq20092wh@yahoo.com.cn 
CAO Ting Oncology Center, Union Hospital, Huazhong University of Science and Technology, Tongji Medical College, Wuhan 430023, China  
LIANG Zhi-wen Oncology Center, Union Hospital, Huazhong University of Science and Technology, Tongji Medical College, Wuhan 430023, China  
CHEN Mi Oncology Center, Union Hospital, Huazhong University of Science and Technology, Tongji Medical College, Wuhan 430023, China  
Hits: 3100
Download times: 2616
Abstract::
      Objective To explore the dose rate response characterization of four kinds of dosimeters for clinical application. Methods Within the range of 100-600 cGy/min, the dose rate responses of the PTW 0.6 cm3 ion chamber, 0.015 cm3 ion chamber, Matrixx Evolution 2D diode array and MapCHECK 2D diode array under the same measuring conditions were measured. The dose rate response of the PTW 0.6 cm3 ion chamber under different energy and working voltage were analyzed.Results All ionization chamber types of measured equipment showed certain dose rate dependence for 6 MV X-rays. All the differences were below 1%.The dose rate dependence disappeared for 15 MV X-rays. The 2D diode array had strong dose rate dependence and the response difference was about 2%.Conclusions It is necessary to test and analyze the dose rate response of the measured equipment in treatment technology with dose rate varying, in order to ensure the precision of daily calibration and dose verification.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9066247  On-line:0

v
Scan QR Code
&et=DB555CA42BE2DFF360BFA576D0F3DFDA49BD4DF8EFEA60CC9E82F2277954B70DADF30CFDC7FDA26CC6F4C64911559977F8C5198857BF17029B0985E6DD198321D3F0937CE7508C145A01A795A8ED08F0DCE0A04C03114FA2C716E23E63C30B1B&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=99E9153A83D4CB11&aid=A239026EAED743DDCE580A6F01EF9CB3&vid=&iid=94C357A881DFC066&sid=F3549E0657848E2A&eid=2A2AA8B7E19F0DF7&fileno=20120523&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="99E9153A83D4CB11"; var my_aid="A239026EAED743DDCE580A6F01EF9CB3";