WU Yun-yun,CUI Hong-xing,ZHANG Qing-zhao,SHANG Bing,SU Xu.Analysis of international intercomparisons results organized by Japan for integrating 222Rn-220Rn detectors[J].Chinese Journal of Radiological Medicine and Protection,2012,32(5):514-517
Analysis of international intercomparisons results organized by Japan for integrating 222Rn-220Rn detectors
Received:March 26, 2012  
DOI:10.3760/cma.j.issn.0254-5098.2012.05.017
KeyWords:222Rn  220Rn  Integrating radon/thoron detectors  International intercomparison
FundProject:国家自然科学基金(10675105)
Author NameAffiliationE-mail
WU Yun-yun Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China  
CUI Hong-xing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China  
ZHANG Qing-zhao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China  
SHANG Bing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China shangbing100088@yahoo.com.cn 
SU Xu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Centre for Disease Control and Prevention, Beijing 100088, China  
Hits: 3943
Download times: 2152
Abstract::
      Objective To guarantee the quality of measurements with the radon-thoron discriminative detectors of our laboratory. Methods LD-P radon-thoron discriminative detector participated in the international intercomparison for integrating radon/thoron detectors organized by National Institute of Radiological Science (NIRS,Japan). Detectors were sent to NIRS for exposure. Radon intercomparison was conducted with radon chamber providing three levels of exposure:low, medium and high levels. Thoron intercomparison was carried out at thoron chamber, which also provided three levels of exposure:low, medium and high levels. Detectors were posted back to our laboratory for etching and analysis after exposure. Then the measured values were submitted to NIRS. Finally the reference values were informed of us.Results The relative percent difference (RPD) between the measured value and the reference value for radon was -13.8%, -14.4% and -17.1% at low, medium and high levels respectively, and that of thoron were -14.4%, 8.9% and -3.2% at three levels respectively.Conclusions Both radon and thoron measurement of our detectors rank as "Category I" in the 4th international intercomparisons for integrating radon/thoron detectors with the NIRS radon/thoron chambers.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9063293  On-line:0

v
Scan QR Code
&et=6FD48CA0A0DD47597E9BF9A9657D9135E45007C2AB112A69D29240BE7FC5BA8B19E0D29DD547A6F28DEE38919B44E1403EC2995C3DC46B2FA11E30A93F860EC2DE2952018180B0DAEFEEFB10B76FB0E87B6DCF3521101BF37BCFEB943DE68A7ECB81B4A5D911AA36DCB2CCA32AFA81F45B67C29FD763B2C5F40629E007AFE9ED8706D7503FA9CA12&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=99E9153A83D4CB11&aid=82F2CF3652F698A61B1D380F5E0526B2&vid=&iid=94C357A881DFC066&sid=C4BBAD7A2DCC89BC&eid=5A751AE9FA58A3FB&fileno=20120517&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="99E9153A83D4CB11"; var my_aid="82F2CF3652F698A61B1D380F5E0526B2";