GAN Jia-ying,HU Yin-xiang,LUO Yuan-qiang,HONG wei,WANG Zhi-yong,LU Bing,JIN Feng.Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose[J].Chinese Journal of Radiological Medicine and Protection,2012,32(4):386-389
Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose
Received:February 26, 2012  
DOI:10.3760/cma.j.issn.0254-5098.2012.04.013
KeyWords:Treatment tabletop  Source-axis distance technology  Dose attenuation  Attenuation factor  Radiotherapy dose
FundProject:贵州省科学技术基金[黔科合(200)号2059]
Author NameAffiliationE-mail
GAN Jia-ying Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China  
HU Yin-xiang Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China  
LUO Yuan-qiang Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China  
HONG wei Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China  
WANG Zhi-yong Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China  
LU Bing Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China lbgymaaaa@sohu.com 
JIN Feng Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China  
Hits: 3313
Download times: 2197
Abstract::
      Objective To investigate the dosimetric influence of pure carbon fiber treatment tabletop of Elekta Precise new linear accelerator in radiotherapy.Methods Surface-axis distance (SAD) technology was employed for the measurement. Two groups of fields were set and both of them were SAD opposed portals (one of them went through the tabletop, while the other did not). A PTW electrometer and a 0.6 cm3 Farmer ionization chamber were utilized for comparison measurement. Then dose attenuation of the main table board, extended body board, the extended board for head, neck and shoulders, and the joints of these boards were calculated. Results Under the energy of 6 MV, the dose attenuations of the following locations were: 1.4%-7.2% at the main treatment table board; 2.8%-38.7%, 1.4%-30.1%, 1.5%-20.8% and 1.4%-11.2%, respectively at distances of 1, 4, 7 and 8 cm from the joint of the main table board;0.5%-5.0% at the extended body board; 4.7%-15.4% at distance of 1cm from the joint of the extended body board; 0.5%-3.3% at the neck position of the extended board for head, neck and shoulders; 5.3%-16.7% at the shoulder positions; and 6.8%-30.4% at the joint between the extended boards and the main table board. Conclusions The dose attenuations of the new linear accelerator pure carbon fiber treatment tabletop vary at different locations. Considerable higher attenuations are observed at the table board joints than other locations.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061939  On-line:0

v
Scan QR Code
&et=FF4EC6B96077DFE7479448A90E6E2A5811EB132629AC0309C89969DC1D0EB49B11FBE98033274BA1D5FFECBD3678105A2519E64890DFB46DF0B16A3111AF716512FF5A2D085798D95191CFF7579AA06127265484F80AD74D639F3DDFB35655153D13D6EE7CCD708B&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=99E9153A83D4CB11&aid=9B3C227A81E2A28CDF4C27FBBCB9BECC&vid=&iid=E158A972A605785F&sid=B54EB62E9D3FF31E&eid=40862C82D74F7473&fileno=20120413&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="99E9153A83D4CB11"; var my_aid="9B3C227A81E2A28CDF4C27FBBCB9BECC";