GAN Jia-ying,HU Yin-xiang,LUO Yuan-qiang,et al.Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose[J].Chinese Journal of Radiological Medicine and Protection,2012,32(4):386-389 |
Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose |
Received:February 26, 2012 |
DOI:10.3760/cma.j.issn.0254-5098.2012.04.013 |
KeyWords:Treatment tabletop Source-axis distance technology Dose attenuation Attenuation factor Radiotherapy dose |
FundProject:贵州省科学技术基金[黔科合(200)号2059] |
Author Name | Affiliation | E-mail | GAN Jia-ying | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | | HU Yin-xiang | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | | LUO Yuan-qiang | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | | HONG wei | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | | WANG Zhi-yong | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | | LU Bing | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | lbgymaaaa@sohu.com | JIN Feng | Department of Oncology, Guiyang Medical College Hospital, Tumor Hospital of Guizhou Province, Guiyang 550004, China | |
|
Hits: 4054 |
Download times: 2617 |
Abstract:: |
Objective To investigate the dosimetric influence of pure carbon fiber treatment tabletop of Elekta Precise new linear accelerator in radiotherapy.Methods Surface-axis distance (SAD) technology was employed for the measurement. Two groups of fields were set and both of them were SAD opposed portals (one of them went through the tabletop, while the other did not). A PTW electrometer and a 0.6 cm3 Farmer ionization chamber were utilized for comparison measurement. Then dose attenuation of the main table board, extended body board, the extended board for head, neck and shoulders, and the joints of these boards were calculated. Results Under the energy of 6 MV, the dose attenuations of the following locations were: 1.4%-7.2% at the main treatment table board; 2.8%-38.7%, 1.4%-30.1%, 1.5%-20.8% and 1.4%-11.2%, respectively at distances of 1, 4, 7 and 8 cm from the joint of the main table board;0.5%-5.0% at the extended body board; 4.7%-15.4% at distance of 1cm from the joint of the extended body board; 0.5%-3.3% at the neck position of the extended board for head, neck and shoulders; 5.3%-16.7% at the shoulder positions; and 6.8%-30.4% at the joint between the extended boards and the main table board. Conclusions The dose attenuations of the new linear accelerator pure carbon fiber treatment tabletop vary at different locations. Considerable higher attenuations are observed at the table board joints than other locations. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|