ZHANG Yi-pei,ZHANG Xue-ning,LIU Dong,LI Yan-he,XU De-sheng.Influencing factors of geometrical accuracy in magnetic resonance imaging for gamma knife radiosurgery[J].Chinese Journal of Radiological Medicine and Protection,2012,32(4):383-385
Influencing factors of geometrical accuracy in magnetic resonance imaging for gamma knife radiosurgery
Received:March 20, 2012  
DOI:10.3760/cma.j.issn.0254-5098.2012.04.012
KeyWords:Gamma knife  Radiosurgery  Magnetic resonance imaging  Geometrical accuracy  Error
FundProject:天津市卫生局科技基金(2010kz92)
Author NameAffiliationE-mail
ZHANG Yi-pei The Second Hospital of Tianjin Medical University, Tianjin 300211,China  
ZHANG Xue-ning 300211 天津医科大学第二医院放射科  
LIU Dong The Second Hospital of Tianjin Medical University, Tianjin 300211,China  
LI Yan-he The Second Hospital of Tianjin Medical University, Tianjin 300211,China  
XU De-sheng The Second Hospital of Tianjin Medical University, Tianjin 300211,China gammaknife@126.com 
Hits: 2951
Download times: 2413
Abstract::
      Objective To evaluate and improve the geometrical accuracy of the magnetic resonance imaging used for gamma knife radiosurgery localization. Methods Four common MRI scan programs were used in this study. The images were transferred to the Gamma Plan treatment planning system. Deviations between stereotactic coordinates based on MR images and real geometrical positions given by the construction of the phantom insert were evaluated for each study. The deviations were further investigated as a function of the spatial position of measured points in the investigated volume.Results The geometrical coordinate errors for axial and coronal images of the selected scan programs were all less than 1 mm. Optimal imaging procedure in terms of geometrical accuracy was observed using spin echo T1 and 3D FSPGR MRI sequences in the unit. Statistically significant differences were found in axial images among SE T1,FSE T2,FRFSE Pd,FSPGR 3D sequences (F=1497.51, 1873.63, 2.87, 3.00,P<0.05). The error increased along with the increase of the distance between node point and centricity. Conclusions The geometrical coordinate errors is related with the imaging program in MRI. Proper imaging programs and parameters should be selected according to the imageology character of patients’ lesions when the radiosurgery is performed.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9063246  On-line:0

v
Scan QR Code
&et=22F1D5E3CC41362BEAB9DA3B61E0EE053E67B04D556C52620FCEB1658870D4954E5087EDBC279FE995F564851108A72E300E8EA16E26C6B04CE4EEA398424C1B32D508615D731E0477B3D1062F560FFC9E67C6B3829743CD4E79AB4E8EF827B43B4EBAB60739DF53&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=99E9153A83D4CB11&aid=9B3C227A81E2A28CEB1B5AB35F433E91&vid=&iid=E158A972A605785F&sid=663FF78B2ADE7A2D&eid=44FDB9366EDDFA2B&fileno=20120412&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="99E9153A83D4CB11"; var my_aid="9B3C227A81E2A28CEB1B5AB35F433E91";