CHI Zi-feng,LIU Dan,CAO Yan-kun,LI Run-xiao,HAN Chun.Pertinence analysis of intensity-modulated radiation therapy dosimetry error and parameters of beams[J].Chinese Journal of Radiological Medicine and Protection,2012,32(3):294-296
Pertinence analysis of intensity-modulated radiation therapy dosimetry error and parameters of beams
Received:November 02, 2010  
DOI:10.3760/cma.j.issn.0254-5098.2012.03.018
KeyWords:Absolute dose verification  Parameters of beams  Verification errors
FundProject:河北省卫生厅科研基金项目(07103)
Author NameAffiliationE-mail
CHI Zi-feng Department of Radiation Oncology, Fourth Hospital of Hebei Medical University, Shijiazhuang 050011, China  
LIU Dan Department of Radiation Oncology, Fourth Hospital of Hebei Medical University, Shijiazhuang 050011, China Liudan2124@sina.com 
CAO Yan-kun Department of Radiation Oncology, Fourth Hospital of Hebei Medical University, Shijiazhuang 050011, China  
LI Run-xiao Department of Radiation Oncology, Fourth Hospital of Hebei Medical University, Shijiazhuang 050011, China  
HAN Chun Department of Radiation Oncology, Fourth Hospital of Hebei Medical University, Shijiazhuang 050011, China  
Hits: 4048
Download times: 2575
Abstract::
      Objective To study the relationship between parameter settings in the intensity-modulated radiation therapy (IMRT) planning in order to explore the effect of parameters on absolute dose verification.Methods Forty-three esophageal carcinoma cases were optimized with Pinnacle 7.6c by experienced physicist using appropriate optimization parameters and dose constraints with a number of iterations to meet the clinical acceptance criteria. The plans were copied to water-phantom, 0.13 cc ion Farmer chamber and DOSE1 dosimeter was used to measure the absolute dose. The statistical data of the parameters of beams for the 43 cases were collected, and the relationships among them were analyzed. The statistical data of the dosimetry error were collected, and comparative analysis was made for the relation between the parameters of beams and ion chamber absolute dose verification results.Results The parameters of beams were correlated among each other. Obvious affiliation existed between the dose accuracy and parameter settings. When the beam segment number of IMRT plan was more than 80, the dose deviation would be greater than 3%; however, if the beam segment number was less than 80, the dose deviation was smaller than 3%. When the beam segment number was more than 100, part of the dose deviation of this plan was greater than 4%. On the contrary, if the beam segment number was less than 100, the dose deviation was smaller than 4% definitely.Conclusions In order to decrease the absolute dose verification error, less beam angles and less beam segments are needed and the beam segment number should be controlled within the range of 80.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9074188  On-line:0

v
Scan QR Code
&et=76848E367C8ACE972FBDAC3737E1E6F38949B420ACA5C37132FE5B047C87A732E90CA023AECC18AF6B1E357A970E42ED26EEFEA0857597119D615B5A9C8962107AAD32543B89747AC0C25D6492CD7DF68C524844D35143E66226500009D5312CD0902FFE6634C1CD&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=99E9153A83D4CB11&aid=669A8D56CCC2B6A113421FE83E6AA574&vid=&iid=38B194292C032A66&sid=E39A3F4E3A67639B&eid=D0182A31A5EB14BA&fileno=20120318&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="99E9153A83D4CB11"; var my_aid="669A8D56CCC2B6A113421FE83E6AA574";