LIU Hai-kuan,GU Nai-gu,GAO Yi-ming,et al.Impact of tomotherapy beam block on the room shielding thickness[J].Chinese Journal of Radiological Medicine and Protection,2011,31(3):340-342 |
Impact of tomotherapy beam block on the room shielding thickness |
Received:June 03, 2010 |
DOI:10.3760/cma.j.issn.0254-5098.2011.03.023 |
KeyWords:Tomotherapy accelerator Shielding Radiation protection |
FundProject: |
Author Name | Affiliation | LIU Hai-kuan | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China | GU Nai-gu | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China | GAO Yi-ming | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China | HUANG Wei-qin | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China | WANG Feng-xian | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China | WANG Li | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China | WU Jin-hai | Institute of Radiation Medicine, Fudan University, Shanghai 200032, China |
|
Hits: 3721 |
Download times: 2808 |
Abstract:: |
Objective To study the calculation of the room shielding thickness of tomotherapy accelerator, a new type of radiotherapy facility, especially the impact of the beam block on the shielding design. Methods According to the relevant standards, combined with the room geometry, the shielding thickness was calculated without the presence of the beam block, considering the primary beam, the scattered beam and leakage. Meanwhile, the shielding thickness was also calculated as comparison with the presence of the beam block, based on the characteristics of tomotherapy facility and its radiation field. Results There was statistical difference between the shielding thicknesses calculated with the presence of the beam block and those without the beam block, to the primary beam direction including the south wall, north wall, the roof and the floor, the shielding thickness were decreased by 95.59%, 63.63%, 80.73% and 51.30%, respectively. Conclusions For the tomotherapy accelerator, the beam block could be of great help to minify the shielding thickness of the room. The radiation field of the tomotherapy facility could be used for the calculation to improve accuracy, and the shielding thickness can also be estimated by subtracting the initial shielding thickness without beam block of the beam block equivalent thickness in the primary beam direction alternatively. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|