YUAN Ji-long,LUO Su-ming,HE Zhi-jian,et al.Development on Monte Carlo methodology with scatter correction factor of afterloading 192Ir source[J].Chinese Journal of Radiological Medicine and Protection,2011,31(1):79-82
Development on Monte Carlo methodology with scatter correction factor of afterloading 192Ir source
Received:October 21, 2009  
DOI:10.3760/cma.j.issn.0254-5098.2011.01.023
KeyWords:Monte Carlo method  192Ir source  Scatter correction factor  Thimble ionization chamber
FundProject:
Author NameAffiliation
YUAN Ji-long National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
LUO Su-ming National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
HE Zhi-jian National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
GAO Lei National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 3463
Download times: 2519
Abstract::
      Objective To facilitate activity measurement by using the thimble ionization chamber in hospitals, to obtain air kerma scatter correction factor of medical afterloading of 192Ir source by developing an available and convenient calculation method. Methods According to International Atomic Energy Agency (IAEA) 1079 Report to calculate the scatter correction factor of 192Ir source, to measure air kerma of 192Ir source with and without lead shield using thimble ionization chamber. Simulation measurement conditions were used to calculate scatter correction factor of 192Ir source and comparison was made between experimental results and literature records. At the same time, the different ionization chamber models were simulated at different room sizes to obtain scattering correction factor of 192Ir source. Results Comparison was made between the simulation scatter correction factors of 192Ir source and experiment by the shadow shield, and the relative deviation was 0.8%. The deviation of the 192Ir activity calculated according to the simulated scatter correction factor and measured by well type ionization chamber was 2.4%. By comparison between the calculated results by using two kinds of spherical ionization chamber and those ones deduced by IAEA 1079 Report, the relative deviations ranged within 0.3%-0.4%. Five different types of thimble ionization chamber and different room sizes were simulated and calculated by MC simulation, with the relative deviation within 3%. Conclusions Monte Carlo simulation method for calculating afterloading 192Ir sources scatter correction factor is feasible, and this method is convenient for use in the thimble chamber for brachytherapy QA work in the hospital.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11565352  On-line:0

v
Scan QR Code
&et=A90EAC7419FB9B22512F4DFDC8FEFBAD0CE8EE80A8FF8BB60724B59A8ED0532E83183C0597D807131ECB295BBB1A0596B8473600A71C3333A67328DB4AE41215EDF02758D8D0DDBC7E9EDE73BA9D727EDDDA034AACD8C4E31CAE55A11152A0A48F2BEC0B33099D6CA3A290275E9E53FD&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9377ED8094509821&aid=44F006F75B3385466BB2F64E5B77F4D4&vid=&iid=CA4FD0336C81A37A&sid=9C65ADEB5990B252&eid=0D0D661F0B316AD5&fileno=20110123&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9377ED8094509821"; var my_aid="44F006F75B3385466BB2F64E5B77F4D4";