WANG Jian-chao,XU Hui,WEI Ke-dao,YUE Bao-rong.Energy dependence of three passive solid detectors on low energy X-ray[J].Chinese Journal of Radiological Medicine and Protection,2010,30(6):735-736
Energy dependence of three passive solid detectors on low energy X-ray
Received:December 17, 2009  
DOI:
KeyWords:Mammography  Solid detector  Energy dependence  Dose measurement
FundProject:科技部科研院所社会公益研究专项(2004BIB1J035)
Author NameAffiliationE-mail
WANG Jian-chao National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
XU Hui National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
WEI Ke-dao National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
YUE Bao-rong National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China bryue@163.com 
Hits: 3558
Download times: 2236
Abstract::
      Objective To assess the feasibility of ESD measurements in the mammography by the comparative study of energy dependence of the three passive solid detectors on low energy X-ray. Methods Two thermoluminescent TL) detectors (LiF(Mg,Ti) and LiF(Mg,Cu,P) and one optically stimulated luminescence (OSL) detector(A12O3:C) were irradiated for the same air kerma by ten kinds of beam quality between 25-40 kV in a Mo/Mo low energy X-ray generator,respectively. Results The energy dependences of TL and OSL were 25% and 11% between 25-40 kV, respectively. Conclusions A considerable potential for OSL detector can be used in routine quality control and ESD measurements in the mammography.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9068462  On-line:0

v
Scan QR Code
&et=AF86A8F1F871ABAD9BE052AA1A1375ACBFE1B694460AB29AD7E9379A19A0E7C5EBD52967F1B1B44AD2177611308C51582FAF61622E6D68A9243FF755B4E1EEA6525E9F63DFE643B3&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=140ECF96957D60B2&aid=4A77CC8A5B70D046C36751AE5655FE03&vid=&iid=B31275AF3241DB2D&sid=B60458D1AE87BCD1&eid=FA98B938D085B826&fileno=20100628&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="140ECF96957D60B2"; var my_aid="4A77CC8A5B70D046C36751AE5655FE03";