WANG Jian-chao,XU Hui,WEI Ke-dao,et al.Energy dependence of three passive solid detectors on low energy X-ray[J].Chinese Journal of Radiological Medicine and Protection,2010,30(6):735-736 |
Energy dependence of three passive solid detectors on low energy X-ray |
Received:December 17, 2009 |
DOI: |
KeyWords:Mammography Solid detector Energy dependence Dose measurement |
FundProject:科技部科研院所社会公益研究专项(2004BIB1J035) |
Author Name | Affiliation | E-mail | WANG Jian-chao | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | XU Hui | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | WEI Ke-dao | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | YUE Bao-rong | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | bryue@163.com |
|
Hits: 4294 |
Download times: 2539 |
Abstract:: |
Objective To assess the feasibility of ESD measurements in the mammography by the comparative study of energy dependence of the three passive solid detectors on low energy X-ray. Methods Two thermoluminescent TL) detectors (LiF(Mg,Ti) and LiF(Mg,Cu,P) and one optically stimulated luminescence (OSL) detector(A12O3:C) were irradiated for the same air kerma by ten kinds of beam quality between 25-40 kV in a Mo/Mo low energy X-ray generator,respectively. Results The energy dependences of TL and OSL were 25% and 11% between 25-40 kV, respectively. Conclusions A considerable potential for OSL detector can be used in routine quality control and ESD measurements in the mammography. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|