MA En-sen,WANG Wu,MA Guo-lin,et al.Accuracy of coronary calcium measurements with low dose 64-slice multi-detector row CT[J].Chinese Journal of Radiological Medicine and Protection,2010,30(5):605-609
Accuracy of coronary calcium measurements with low dose 64-slice multi-detector row CT
Received:June 02, 2010  
DOI:
KeyWords:Coronary artery disease  Tomography, X-ray computed  Atherosclerosis  Calcium
FundProject:
Author NameAffiliationE-mail
MA En-sen Department of Radiology, China-Japan Friendship Hospital, Beijing 100029, China Mgl4022005@yahoo.com.cn 
WANG Wu Department of Radiology, China-Japan Friendship Hospital, Beijing 100029, China  
MA Guo-lin Department of Radiology, China-Japan Friendship Hospital, Beijing 100029, China  
YANG Zhi-gang Department of Radiology, China-Japan Friendship Hospital, Beijing 100029, China  
LUAN Hui 四川大学华西医院放射科  
Hits: 3522
Download times: 3227
Abstract::
      Objective To compare the low- and standard-dose protocols in the evaluation of coronary artery calcification (CAC) scores and in image quality, and to assess the accuracy of low-dose prospective ECG-gated helical CT in CAC measurements. Methods Forty-three patients with CAC were scanned twice using a tube current setting of 100 and 55 mA retrospectively. CAC scores (Agatston) of different branches of coronary artery were recorded. The mean and SD of the CT attenuation values in regions of interest in the aorta were measured, and the value (mean+2×SD) were obtained. Results Calcification plaques were found in 105 of the 172 vessels in the 43 subjects. The radiation dose of the 50 mAs group was (0.72±0.04) mSv, significantly lower by 0.6 mSv than that of the 100 mAs group [(1.32±0.08)mSv, P <0.05]. The total calcification score of the 55 mAs group was significantly positively correlated with the routine dose group after performing the square root transformation (r =0.998). A signficant correlation existed in vessel-based calcification between these 2 groups (r = 0.997). Conclusions Low-dose prospective ECG-gated 64-slice CT yields high accuracy on CAC scoring while maintaining the image quality needed to measure CAC.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11565301  On-line:0

v
Scan QR Code
&et=DCD16A6798AF6C329D081BFDF7A3AFFA7FBD70E67DFC55A8E3AE73C9DDCF9B73926F358A1824FB84E7FDEC37345386F16DFAE2D1F1078CC55F687A4855B9A6DDF2FDCF574100E77B7B53A3D772D51CE24DEBD3B10B0B14E0&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=140ECF96957D60B2&aid=30052B74ACA90FDF591D3367CE8D9E4E&vid=&iid=94C357A881DFC066&sid=10A39635766FF5D0&eid=5568599C60D4BE87&fileno=20100531&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="140ECF96957D60B2"; var my_aid="30052B74ACA90FDF591D3367CE8D9E4E";