LU Xiao-guang,WANG Yun-lai,HUO Xiao-qing,SHA Xiang-yan,LIAO Xiong-fei.Penumbra measuremnts of BeamModulatorTM multileaf collimator[J].Chinese Journal of Radiological Medicine and Protection,2010,30(4):452-455
Penumbra measuremnts of BeamModulatorTM multileaf collimator
Received:October 04, 2009  
DOI:10.3760/cma.j.issn.0254-5098.2010.04.023
KeyWords:Radiotheray  Multileaf collimator  Penumbra  Dosimetry
FundProject:
Author NameAffiliationE-mail
LU Xiao-guang Department of Radiation Oncology, PLA General Hospital, Beijing 100853, China  
WANG Yun-lai Department of Radiation Oncology, PLA General Hospital, Beijing 100853, China nanyangwang@163.com 
HUO Xiao-qing Department of Radiation Oncology, PLA General Hospital, Beijing 100853, China  
SHA Xiang-yan Department of Radiation Oncology, PLA General Hospital, Beijing 100853, China  
LIAO Xiong-fei Department of Radiation Oncology, PLA General Hospital, Beijing 100853, China  
Hits: 3956
Download times: 2310
Abstract::
      Objective To evaluate the penumbra of a new multileaf collimator equipped with Elekta Synergy accelerator.Methods The penumbra were derived from beam profiles measured in air and water using PinPoint ion chamber with PTW MP3 water phantom. Variations of penumbra with X-ray beam energy, depth in water, and leaf position were investigated.Results The penumbra in air for 6 MV X-ray was 2 mm less than that at depth of maximal dose in water. The penumbra of leaf side was 1 mm less than that of the leaf end. The penumbra had close relationship with beam energy, depth in water and leaf position. Penumbra was increased with beam quality and water depth. The leaf position had great influence on the penumbra.Conclusions The penumbra of the multileaf collimator is related to its original design and radiation delivery technique. Special considerations should be taken into during treatment planning. Regular measurement should be performed to guarantee the delivery quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9074445  On-line:0

v
Scan QR Code
&et=7EA0FCDEF2CAA2747B5FE0151CA217716FCC433F5315E0E64BFC6B6BAD34AED5BED2C23A3C683D2F37C1D425B10A5EF41A40B615539BB901AE913FC8DF41B956BD416DA542EDABBC&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=140ECF96957D60B2&aid=736D0131EBC47E21429B60E3B0D58EDC&vid=&iid=E158A972A605785F&sid=30F3EEEA29E34EE7&eid=366B1248A15658C5&fileno=20100423&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="140ECF96957D60B2"; var my_aid="736D0131EBC47E21429B60E3B0D58EDC";