CAO Lei,DENG Jun,WANG Cheng-guo,et al.Design to neutron shielding for medical LINAC therapy room[J].Chinese Journal of Radiological Medicine and Protection,2009,29(5):510-512
Design to neutron shielding for medical LINAC therapy room
Received:September 03, 2008  
DOI:
KeyWords:Medical accelerator  Neutron contamination  Scattering in maze  Dose estimation
FundProject:
Author NameAffiliation
CAO Lei National Institute for Radiological Protection and Nuclear Safety, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
DENG Jun National Institute for Radiological Protection and Nuclear Safety, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
WANG Cheng-guo 内蒙古疾病预防控制中心 
JIAO Ling 中国医学科学院放射医学研究所 
ZHANG Wei-jun National Institute for Radiological Protection and Nuclear Safety, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
ZHAO Lan-cai National Institute for Radiological Protection and Nuclear Safety, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 3915
Download times: 2896
Abstract::
      Objective To study the source of neutron in the therapy room and the essence of neutron scattering, to analyze the change patten of neutron dose in the therapy room as well as in the maze so as to design the shielding. Methods Based on the measurement of the neutron flux for concerning points at the patient plane caused by a running 15MeV accelerator and referred to NCRP report 79, this paper carried out the radiation protection design and calculation. Results When X-ray produced by medical accelerator reached certain energy, photonuclear reaction is the main source of neutron contamination in medical accelerator room. The main target of protection is the scattered neutron that comes to inside entrance of the maze and the γ-ray caused by them. As a result neutron contamination has nothing to do with the therapeutic effect, but only to increase the dose commitment for relevant person. Conclusions Neutron contamination has nothing to do with therapeutic effect, but only to increase the dose commitment for relevant person. Under certain conditions, it may also cause radiation insult to them. Therefore, certain attention should be paid the hazard caused by neutron external exposure, at the same time shielding design and evaluation should be implemented against the neutron contamination in medical accelerator room.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11566204  On-line:0

v
Scan QR Code
&et=045035D11597C08A29302DA204BBF6AACCB289B6132172DE11A398698F097CCEF4210CF06F61C4472401AAE069B79DAE57F8ADE8560E34E267D4BF3A914F7645&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=DE12191FBD62783C&aid=EBFE44D24D47BBCF6F269DB27C65F638&vid=&iid=94C357A881DFC066&sid=25467A5A28500A25&eid=D8AE57480552698F&fileno=20090515&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="DE12191FBD62783C"; var my_aid="EBFE44D24D47BBCF6F269DB27C65F638";