LI Heng-xiang,ZHANG Hua-fang,CHEN Pei-zhong,et al.Evaluation of angle regulator with protection shields in the skeletal fluorosis X-ray radiography[J].Chinese Journal of Radiological Medicine and Protection,2009,29(4):423-425 |
Evaluation of angle regulator with protection shields in the skeletal fluorosis X-ray radiography |
Received:May 18, 2009 |
DOI: |
KeyWords:Skeletal fluorosis Forearm Shank Angle regulator X-ray protection |
FundProject:山东省医药卫生科研基金(2004HZ086) |
Author Name | Affiliation | LI Heng-xiang | Shandong Institute for Prevention and Treatment of Endemic Diseases, Jinan 250014, China | ZHANG Hua-fang | Shandong Institute for Prevention and Treatment of Endemic Diseases, Jinan 250014, China | CHEN Pei-zhong | Shandong Institute for Prevention and Treatment of Endemic Diseases, Jinan 250014, China | ZHANG Hai-yan | Shandong Institute for Prevention and Treatment of Endemic Diseases, Jinan 250014, China | YANG Fang | Shandong Institute for Prevention and Treatment of Endemic Diseases, Jinan 250014, China |
|
Hits: 3334 |
Download times: 2126 |
Abstract:: |
Objective To evaluate the value of the multifunctional angle regulator with shields in skeletal fluorosis X-ray radiography. Methods Conventional radiography, regulator with shields and regulator for radiography were used to examine the forearms and shanks of 78 patients, and then the quality of the photos and the diagnostic results were compared. The dose rates on patients' foresides, sides, and bottoms of division 1 and 2 m were measured by pressure ionization instrument. Results Regulator with shields and regulator for radiography had a higher rate of good quality than conventional radiography by 39%, and consequently the diagnostic accuracy rate were improved by 37%(χ2=43.33,43.98 and 17.76,14.56,P<0.01). By conventional radiography the average dose rates on forearms and shanks 1 and 2 m away from X-ray axis were 1198, 934 and 1439, 1175 μSv/h, respectively; while regulator with shields and single regulator were used, corresponding average dose rates were 80.6,39.4 and 95.6,59.5 μSv/h and 375.7,187 and 447,226.3μSv/h,respectively. Conclusions Angle regulator with shields is effective devices to examine the forearms and shanks of the patients who are suffering from dysfunction of skeletal fluorosis. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|