Scatter filtration method in digital radiography: The imaging quality and irradiation dose evaluation for scatter filtering template performance
Received:February 28, 2008  
DOI:10.3760/cma.j.issn.0254-5098.2009.02.041
KeyWords:Digital radiography  Scatter  Imaging quality  Radiation dose  Grid
FundProject:
Author NameAffiliation
WANG Peng-cheng Department of Radiology, Taishan Medical College, Taian 271016, China 
ZHANG Gui-fang 山东省肿瘤医院 
HOU Li-xia Department of Radiology, Taishan Medical College, Taian 271016, China 
谢晋东 Department of Radiology, Taishan Medical College, Taian 271016, China 
李士骏 苏州大学放射医学与公共卫生学院 
Hits: 5364
Download times: 3188
Abstract::
      Objective The imaging quality and irradiation dose were compared between scatter filtering template performance and conventional grid in digital radiography.Methods Images of contrast-detail phantom and ROC statistic phantom corrected with scatter filtering template method and grid were obtained, respectively. The image quality factors (IQF) and signal detection probability values (Pdet) and imaging dose were analyzed by ANOVA, SNK and t test statistics for the two methods of scatter removing.Results The difference of IQF and signal detection probability of images corrected by scatter filtering template technique and conventional grid was statistically significant.Conclusion In digital radiography, STFF can effectively filter scattering, improve image quality without increasing the exposure. With the same image quality, it can reduce the entrance surface dose by about 30% than filter grid. In order to effectively filter scattering, improve the quality of X-ray image, and reduce the patient dose, STFF is an alternative method.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11566867  On-line:0

v
Scan QR Code
&et=DFC906FA2B5504331D4FE13F8E938410A9CD76CD7AE684B57DA02F5B0EE1A331D5AD8D7595980D1B5846C502B56A222490C49C01945BD81071D8D1EFC4813C35B124FFAE5F463EB7F504FFE34003E42E8C1C18D468C46E8E2AD452B964515D1D7E91AB67969B6451E115B185E22513CCFBB679CD1380EF077F5179A64C68C07558E9EB39402EE38896BDF55EEC2D8D0FDD50D451F3A13088&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=DE12191FBD62783C&aid=1D1F68D2FB5F06A40930819878A832E1&vid=&iid=0B39A22176CE99FB&sid=D6354F61445E9456&eid=E1D946F217E3B046&fileno=20090241&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="DE12191FBD62783C"; var my_aid="1D1F68D2FB5F06A40930819878A832E1";