CHEN Shao-wen,TANG Qiang,LIU Xiao-wei.Absorbed dose response of Al2O3 dosimeter irradiated by 60Co γ spectrum source capture and collimators[J].Chinese Journal of Radiological Medicine and Protection,2009,29(1):91-93 |
Absorbed dose response of Al2O3 dosimeter irradiated by 60Co γ spectrum source capture and collimators |
Received:February 26, 2008 |
DOI:10.3760/cma.j.issn.0254-5098.2009.01.030 |
KeyWords:Monte Carlo simulations Al2O3 dosimeter Absorbed dose response EGSnrc |
FundProject: |
|
Hits: 3518 |
Download times: 1901 |
Abstract:: |
Objective To investigated the absorbed dose response of Al2O3 dosimeter in water phantom irradiated by 60Co γ spectrum source.Methods The EGSnrc simulation program code DOSRZnrc was used to calculate the absorbed dose of the Al2O3 dosimeter and that of the equivalent volume of water in the corresponding position, as well as the absorbed dose conversion factor, irradiated by 60Co photon beams in a water phantom. Simulations were done for a cylindrical geometry dosimeter (diameter 0.4 cm and height 0.1 cm) and the dosimeter was placed at the centre of the water phantom at different depths.Results The average absorbed dose conversion factor is 1.143±0.006 and changes little with the depth of the dosimeter in the water phantom, and the deviation is less than 1.0%. Conclusion The absorbed dose response of Al2O3 dosimeter irradiated by 60Co γ spectrum source is steady and is independent on the depth of the dosimeter in water phantom in this research. |
HTML View Full Text View/Add Comment Download reader |