WANG Jian-chao,YUE Bao-rong,WEI Ke-dao.Comparison of the energy response for several dosimeters used in mammography[J].Chinese Journal of Radiological Medicine and Protection,2008,28(2):188-190 |
Comparison of the energy response for several dosimeters used in mammography |
Received:December 23, 2008 |
DOI: |
KeyWords:Mammography dosimeter Energy response Standard radiation quality |
FundProject:科技部科研院所社会公益研究专项资助项目(2004BIB1J035) |
|
Hits: 3900 |
Download times: 2251 |
Abstract:: |
Objective To evaluate whether eight kinds of dosimeters satisfy the clinical mammography dose measurement through the comparison of energy response.Methods According to GB/T 19629-2005,the energy responses of eight available mammography dosimeters were determined for ten standard radiation qualities in the range 25-40 kVp at the SSDL using a refrence-class dosimeter.Results Except 2000S, the energy response factors of all the other dosimeters were consistent within ±5%.Conclusions The energy response of ionizing chamber type is better than other types of dosimeters. |
HTML View Full Text View/Add Comment Download reader |
Close |