WANG Peng-cheng,ZHAO Lei,ZHAI Xiao-juan,et al.Effect of kVp and mAs settings on patient exposure dose and the detecting sensitivity for low contrast blood vessel in digital subtraction angiography[J].Chinese Journal of Radiological Medicine and Protection,2007,27(4):400-402
Effect of kVp and mAs settings on patient exposure dose and the detecting sensitivity for low contrast blood vessel in digital subtraction angiography
Received:July 10, 2006  
DOI:
KeyWords:Digital radiography  Flat panel detector  Exposure dose
FundProject:
Author NameAffiliation
WANG Peng-cheng Department of Radiology,Taishan Medical College,Taian 271016,China 
ZHAO Lei Department of Radiology,Taishan Medical College,Taian 271016,China 
ZHAI Xiao-juan Department of Radiology,Taishan Medical College,Taian 271016,China 
刘伟 Department of Radiology,Taishan Medical College,Taian 271016,China 
张桂芳 苏州大学放射医学与公共卫生学院 
李士骏 苏州大学放射医学与公共卫生学院 
黄永杰 苏州大学放射医学与公共卫生学院 
Hits: 4148
Download times: 2920
Abstract::
      Objective To investigate the effectiveness of kVp, mAs on patient exposure dose and the detecting sensitivity for low contrast blood vessel in digital subtraction angiography. Methods A DSA phantom with 1.5 mm diameter vessels containing iodine at various concentrations was made. The air kerma 1.1 μGy/frame was kept at the entrance of I.I, with kVp changed to measure ESD and to determine the detectable threshold iodine concentration (DTIC) of the phantom on image displayer.X-ray tube voltage was kept at 70 kV, adjusted mAs and measured air kerma at entrance of I.I, ESD and DTIC. The effect dose was calculated with Monte Carlo simulation program PCXMC Ⅴ 1.5. Results As kVp varied from 50 to 100 kV,while the entrance air kerma of I.I was kept at 1.1 μGy/frame, the ESD and effect dose of patient varied linearly according to mAs. As mAs changed from 6.4 to 64 less than 70 kV, the DTIC decreased exponentially, while as the exposure exceed 40 mAs,the improvement of DTIC become modest. Conclusions Reduction of the X-ray tube voltage instead of increase of the exposure level would improve the DSA system performance. As the entrance air kerma of I.I exceed 2.5μGy/frame,increasing exposure may not improve the image quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029817  On-line:0

v
Scan QR Code
&et=FF4EC6B96077DFE7312EA83CD81A79165646C41606F83E32A1180862E0D18FEC4DFA5FFE5EF0955ADB963936BDB09C8314C0407F69BE7C125392FC6820B9245563C109C552C4FCE637D813DBC62F1CA715BF3DEABD99A2A05BE8B4C802D641A50F04086ECE1316D0C1EBD228C220A09B44A1C52F81201BA46154CFFC37AE61BD52E0C0C667BF5FB1EEDB1036C2165040F965886B5980F628&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=A732AF04DDA03BB3&aid=308A5963270DDEEA1ED0B21C621FA87F&vid=&iid=E158A972A605785F&sid=8ACD9060100C26F1&eid=EE23223CE4332BD7&fileno=20070430&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="A732AF04DDA03BB3"; var my_aid="308A5963270DDEEA1ED0B21C621FA87F";