HE Zhi-jian,LUO Su-ming,ZHU Wei-guo,et al.TLD comparisons for radiotherapy dosimetry in non-reference conditions[J].Chinese Journal of Radiological Medicine and Protection,2007,27(3):285-287
TLD comparisons for radiotherapy dosimetry in non-reference conditions
Received:May 30, 2006  
DOI:
KeyWords:Thermoluminescence dosimetry  Non_reference condition  Photon beam  Electron beam
FundProject:国际原子能机构基金资助项目(11916R0)
Author NameAffiliation
HE Zhi-jian National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
LUO Su-ming National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
ZHU Wei-guo National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
李开宝 National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 3543
Download times: 2076
Abstract::
      Objective To carry out TLD comparisons for radiotherapy dosimetry in non-reference conditions in some advanced hospitals in Beijing. Methods The TLD measuring system was calibrated in Co-60 γ-ray beam. The readings were corrected by energy response and TLD holder attenuation correction factors. Results For high energy photon beams, the mean ratio of EAG's dose to user's dose (DEAG/DUSER) is 1.001; For high energy electron beams, the mean ratio of DEAG/ DUSER is 1.000. Conclusions According to IAEA requirement, TLD dose comparison error for hospital, the expanded uncertainty (k=2) shall not be more than 5%. The results indicate that the qualified rates are 98% for high energy photon beams and 100% for high energy electron beams in the participating hospitals.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11566285  On-line:0

v
Scan QR Code
&et=10EB440D897DCD6C10409811908AE9BE0399BEB922446E25F77D0A0FB11DD15C60BA87BDF745A45894E1621EBAB779F46B19707FF515A2730F770CBB149F81A3CD5F4D2DCB282223&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=A732AF04DDA03BB3&aid=8F993DD978B84741D9B8E6BB960F4BCD&vid=&iid=38B194292C032A66&sid=B8F8200D88DDC7D6&eid=EBD6B792C9111B87&fileno=20070328&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="A732AF04DDA03BB3"; var my_aid="8F993DD978B84741D9B8E6BB960F4BCD";